5
High Voltage Probe HV-P60A
3.2 Adjustment
This instrument is connected and adjusted according to the following
procedures.
(1) As shown in Figure 3(a), the ground terminal of the probe body is
connected with the terminal GND of the signal generator.
(2)
INPUT of the oscilloscope is connected with the connector of the
probe body with Output cable.
(3) The output of the signal generator is connected with the tip of
the probe body
Signal Generator
Figure 3 (a) Connection when probe is adjusted
Oscilloscope
Probe body
Ground terminal
Matching Box
Connector
Tip of Probe
Figure 3
(b) Waveform adjustment of probe
in matching box (closeup)
Screwdriver
Hole for waveform adjustment
Clamp
Ou
tp
u
t ca
b
le
Ground lead