Capter2
Name
and Function of Each Part (Outline)
2-14
Curve Tracer CS-5000 Series Instruction Manual
CAUTION
!
About the caution marks and the voltages that display in the panel and the
patch panel of the test fixture
Missing proper wiring and/or wrong connection may conduct wrong measurement. Also these
connections may cause of damage on DUT and/or malfunction of the unit.
Case1: Wrong connection between Collector terminal and Base terminal;
DUT may be damaged while extra high voltage can be supplied on Base terminal.
Case2: NOT connecting Collector output terminal, Base output terminal and/or Emitter terminal
and connecting Collector sense terminal, Base sense terminal and/or Emitter sense terminal
respectively;
It may cause for burn of SENSE RESISTERs inside of the unit since the sense resisters
connected to outputs respectively for Kelvin sensing and may be supplied current exceeding
each specification.
Case3: Connecting HIGH VOLTAGE terminal and HIGH CURRENT terminal of collector output at
the same time on DUT;
It may be cause for fire and/or malfunction due to high voltage applied to HIGH CURRENT
terminal.
2.5 Test fixture CS-304
Figure 2.14 shows the patch panel part on the test fixture CS-304 of this instrument and Table 2.11
on next page lists name and function of each part. CS-304 is the test fixture for only CS-5400.
Figure 2.14 Name and function of each part on test fixture CS-304
⑰
Содержание CS-5000 Series
Страница 1: ...Instruction Manual Curve Tracer CS 5000 series CS 5100 CS 5200 CS 5300 CS 5400 ...
Страница 2: ...Ⓒ 2013 IWATSU ELECTRIC CO LTD All rights reserved ...
Страница 16: ...0 4 Memo ...
Страница 48: ...Capter2 Name and Function of Each Part Outline 2 30 Curve Tracer CS 5000 Series Instruction Manual Memo ...
Страница 210: ...Capter4 Characteristics Measurement 4 40 Curve Tracer CS 5000 Series Instruction Manual Memo ...
Страница 214: ...Capter5 Daily Maintenance and Calibration 5 4 Curve Tracer CS 5000 Series Instruction Manual Memo ...