English
English
English
English
7
Table 2: Error codes
Error code(s) Description
Possible causes and solution(s)
1, 2 , 3
Internal battery charger
problem.
Restart battery charger. If the error is repeated
Consult Your supplier.
4
Temperature too low.
Move to a warmer environment and restart the
charging.
5
Temperature too high
Let the charger cool down for 15 minutes and
restart the charging. If the problem persists
please consult Your supplier.
6
Too many ampere
hours charged.
1: Connected a battery with higher capacity then
specified.
2: Unknown problem. Please consult Your
supplier.
8
Voltage increase
insufficient.
Possibly a defective battery.
Please consult Your supplier.
If the cause of the failure has been removed, the battery charger can be restarted by
disconnecting and reconnecting the mains .
6. Specifications
c
-Go
specifications / model
6A
8A
12A
Supported batteries
Lead Acid (Gel/ AGM) 24V or 2 x 12V
Battery capacity range
40-60Ah
60-85Ah
80-125Ah
Mains voltage
220-240Vac
nominal,
(single phase)
Mains frequency
50/60 Hz
Output voltage range
24V nominal
Output current range
0.25 – 6A
0.25 – 8A
0.25 – 12A
Maximum output power *
180W
240W
360W
Efficiency
> 90% at full load and 230Vac
Protections
Polarity , Output voltage , Temperature
Dimensions
210 x 175 x 65mm
Charge status indication
2 LEDs
Use
Indoor use only
Operational temperature range *
0 – 40°C
Storage temperature
-15 - +50°C
Cooling
Passive
Passive
Active (Fan)
Maximum humidity
95% (non condensing)
Safety class
II
Regulations
CE (LVD, EMC, RoHS)
Standards
EN60335-2-29 , EN12184 , ISO7176-14, EN60601-1-2
* At elevated ambient temperatures or if there is no sufficient cooling, the output current
can be reduced.