26
7.2 Monitoring and maintenance
The time for these maintenance procedures depends on the
frequency of use and the working habits of the users. For that
reason, the recommended times are only approximations.
This furnace has been developed for typical use in dental
laboratories. If the product is used in a production enter-
prise, for industrial applications and for continuous use,
premature ageing of the expendable parts has to be expected.
The expendable parts are as follows:
– Heating muffle
– Insulation material
Expendable parts are not covered by the warranty. Please also
observe the shorter service and maintenance intervals.
If the furnace head is replaced with a furnace head from
another Programat S1, calibration is required.
7.3 Cleaning
The furnace may only be cleaned when it is cool, since
there is a burn hazard. Do not use any cleaning solutions.
Disconnect the furnace from the power source before
cleaning.
The following parts have to be cleaned from time to time:
Item:
Frequency:
Cleaning Material:
Furnace housing (7) and
furnace head
if required
soft, dry cloth
Membrane-sealed
keypad (8)
weekly
soft, dry cloth
Cooling tray (26)
daily
cleaning brush*
Isolation (1)
daily
cleaning brush*
*Never use compressed air!
What:
Measure:
When:
Check all plug-in connections for correct fit
Var. external connections
weekly
Check if the furnace head opens smoothly and without excessive noise
Opening mechanism
monthly
Check if the thermocouple is straight and in the right place
Thermocouple (2)
weekly
Check the insulation for cracks and damages. If the insulation is worn down
it has to be replaced by a certified Ivoclar Vivadent Service Centre. Fine hair-
line cracks on the surface of the insulation are harmless and do not
influence the function of the furnace in a negative fashion.
Insulation (1)
monthly
Check the keypad for visible damage. If the keypad is damaged, it has to be
replaced by a certified Ivoclar Vivadent Service Centre.
Keypad (8)
weekly
Check temperature. Use the temperature checking set to check and adjust
the temperature in the furnace.
Firing chamber
twice a year