20
5.6 Settings / test program and information
By pressing the Settings key (78) you will reach the setting screen
(the latest setting selected will be shown).
The cursor keys (79, 80) can be used to toggle between the possible
settings. This screen can be left by pressing ESC (71) or by pressing
one of the program keys (82, 83, …).
5.6.1 Settings / test programs
Settings
Display
Short Description
Contrast
The contrast can be
adjusted with the -/+
keys.
Temperature unit
The -/+ keys are used
to toggle between °C
and °F.
Language selection
Enables the language
selection.
Temperature
calibration
With this program,
the temperature
calibration program
can be started. See
Chapter 7.5
Temperature Calibra-
tion for more details.
Volume
The desired volume
can be set using the
-/+ keys
Tune
The desired tune may
be set using the -/+
keys
Programming
Enables the pro-
gramming of the
parameters for the
currently selected
program *
Renaming
Enables the renaming
of the currently
selected program *
Enables the renaming
of the material
Time
The time can be set
using the -/+ keys
Settings
Display
Short Description
Date
The date can be set
using the -/+ keys
General write
protection
Enables the activation
or deactivation of the
general write protec-
tion by means of the
-/+ keys after enter-
ing the user code.
The write protection
covers all programs
Heater test program
Enables a check of
the heating system
Keypad test
Enables a check of
the membrane-sealed
keypad
Cleaning program
With this program,
the heating muffle,
including the insu-
la-tion materials, can
be cleaned by means
of a heat process
Protocol
Enables the creation
of protocols of the
sintering procedures
Start delay
Activating this func-
tion enables the
delayed start of
sintering programs
Operating Status
Display
With this menu item,
the Operating Status
Display can be
activated/deactivated
Service interval
Select the interval for
the next reminder.
The intervals are set
in months.
Load factory settings
With this setting
option, all the values
and parameters can
be reset to the
original factory
settings.
Note:
All the
individual programs
that you have
generated and
saved will be deleted
by this function
Important information
For certain settings, the user code (359) is required.
* Some programs (P1–P7) are protected with a code. If changes are necessary, the