![Ivoclar Vivadent bluephase C8 Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/ivoclar-vivadent/bluephase-c8/bluephase-c8_instructions-for-use-manual_2097626051.webp)
55
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 test level
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance level
10 V
10 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should not be used closer to any
part of the Bluephase C8, including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 0.35 P
d = 0.35 P 80 MHz to 800 MHz
d = 0.70 P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
Guidance and manufacturer‘s declaration - electromagnetic immunity
The Bluephase C8 is intended for use in the electromagnetic environment specified below. The customer or the user of the Bluephase C8 should assure
that it is used in such an environment.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be con-
sidered. If the measured field strength in the location in which the Bluephase C8 is used
exceeds the applicable RF compliance level above, the Bluephase C8 should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the Bluephase C8.
b
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the Bluephase C8
The Bluephase C8 is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the Bluephase C8 can help prevent electromagnetic interference by maintaining a miminum distance between portable and
mobile RF communications equipment (transmitters) and the Bluephase C8 as recommended below,
according to the maximum output power of the communication equipment.
Rated
maximum
output power
of transmitter
W
0.01
0.1
1
10
100
150 kHz to
80 MHz
d = 0.4 P
0.04
0.13
0.40
1.3
4.0
Separation distance according to
frequency of transmitter
m
80 MHz to
800 MHz
d = 0.4 P
0.04
0.13
0.40
1.3
4.0
800 MHz to
2.5 GHz
d = 0.7 P
0.07
0.22
0.70
2.2
7.0
Содержание bluephase C8
Страница 27: ...31 o oo 32 32 32 o o o o 34 o 34 o o o o o o o o o o o 36 o o o 36 o 38 o 38 o 38 39 39 39 40 40 o o o 41...
Страница 30: ...34 o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o...
Страница 31: ...35 o o o o o o o o o o o o o o o o o o o o o o o o...
Страница 38: ...42 Bluephase C8 Ivoclar Vivadent...
Страница 39: ...43 44 44 44 46 46 48 Cure Memory 48 50 50 50 51 51 51 52 52 53...
Страница 41: ...45 515 40 20...
Страница 42: ...46...
Страница 43: ...47...
Страница 46: ...21 Cure Memory 10 800 2 10 10 10 8 Pin Point 6 2 Pin Point Bluephase C8 10 Bluephase Meter 50...
Страница 47: ...22 FD366 D rr Dental Incidin Liquid Ecolab 40 10 pH 5 5 pH 8 5 10 51...
Страница 48: ...23 Bluephase C8 3 Vivadent Ivoclar Ivoclar Vivadent 52...
Страница 52: ...56...
Страница 53: ...57...