6
SETUP PROCEDURE
CAUTION:
Do not use any kind of extender cables between oral
or rectal probe cables and instrument. Instrument will not operate if
probes are misconnected.
7. a. Model 2180 - Attach carrying strap to thermometer, if
desired. Place the thermometer in the Home Base to
activate. Always store the instrument in the home base
when not in use.
b. Model 2185 - Remove the two hex drive screws on both
sides of the Home Base using the tool provided. Place the
thermometer in the Home Base to activate. Tighten the
screws using the tool provided.
CAUTION:
Do not over tighten the screws.
1. Place the TURBO TEMP™ Electronic Thermometer in the
home base before using it for the first time or the instrument
will not operate. The instrument must also be returned to
the home base whenever the batteries are replaced.
2. The anti-theft timer and alarm are reset by returning the
thermometer to the home base. When the anti-theft timer is
configured for 8 hour operation (FACTORY DEFAULT), the
RETURN TO BASE
message will be displayed approximately
one hour before the timer interval expires. When the anti-
theft timer is configured for one hour operation, the
RETURN TO BASE
message will be displayed approximately
10 minutes before the timer interval expires. When the
timer interval expires (either 8 hours or 1 hour) the
instrument will alarm with 3 long beeps or continuous
beeps, (depending upon configuration settings) flash
RETURN TO BASE
message and cease to function.
Home Base
INST
ALLA
TION
*
Содержание 218X
Страница 1: ...TURBO TEMP ELECTRONIC THERMOMETER MODELS 218X DIRECTIONS FOR USE ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 12: ......
Страница 20: ...16 FEATURES AND OPTIONS ACCESSORIES Accessories 2031 ...