FEATURES AND OPTIONS
13
LOW BATTERY
- Indicates batteries are low. Battery replacement
is not required until the
REPLACE BATTERY
alarm is displayed
(see below).
BATTERY REPLACE
- Indicates the batteries are unusable and
must be replaced.
RETURN TO BASE
- Displayed continuously indicates that 10
minutes (for 1 hour configuration) or 1 hour (for an 8 hour
configuration) or less remains before the expiration of the
predetermined anti-theft time period (factory set to 8 hours).
When the configured time period expires, the instrument will
emit 3 long or continuous beeps (depending upon configuration
settings), display the
RETURN TO BASE
message and cease to
function. Return to home base to reset anti-theft timer.
BROKEN PROBE
- (accompanied by
FIX M E
and broken probe
indicator, either oral or rectal) - Indicates a probe is broken and
needs to be replaced. Replace probe in well or unplug probe to
reset. Replace with new probe if the message persists.
FIX ME #
(accompanied by three long beeps and a number) -
Indicates instrument detects a self-test failure. Instrument
operation is discontinued until the failure is corrected. Remove
one battery, then reinstall battery. The instrument will run
through the self-test routine. If the fix me message appears
again, record the fix me number and refer to the Service
Information section of this manual.
NORMAL USE
TISSUE CONTACT PINWHEEL:
Three arrows, sequentially forming a circle, with
arrowheads rotating clockwise. Motion indicates
tissue contact.
PULSE TIMER CLOCK:
Consists of twelve pie wedges forming a clock.
Provided to time events such as pulse rate or
respiration rate.
Figure 10 - Thermometer Display