These ThermoProbe instruments are intended for use in both hazardous (potentially flammable or explosive) and non-
hazardous areas under dry conditions at ambient temperatures between -20

°

C to 40

°

C.  

The instruments are not intended for use in permanent outdoor installations and are not intended or tested for icing 
conditions. Additional means of protection should be used where the equipment may be exposed to excessive external 
stresses (e.g. vibration, heat, impact, etc.).

 

The user must have a working knowledge of appropriate safety 

requirements. Be aware of aggressive substances and that extra protection may be needed. Exposure to excessive heat 
can melt the plastic components of the instrument. 
 
a) The user must have a thorough knowledge of the products to be measured and must know of the safety precautions 
to be taken when working with the material to be measured.  
b) The instrument shall be checked concerning severe defects; check that instrument is complete, has good batteries, 
etc. If necessary, check measurement accuracy. If any defects are found, the instrument should not be used until 
repairs have been made.  
c) The instrument should be clean for safety and ease of use.  
d) The physical measurement location should be evaluated for primary and secondary risks.  
e) Power source must be removed before performing any maintenance.   
f) Exchange of components other than the batteries may compromise ATEX/IECEx or other certifications and shall 
only be undertaken by ThermoProbe or one of its qualified service providers. See also “Authorized Repair” section.  
 
CAUTION:  
The following metal parts have a capacitance in excess of 3pF, are isolated from ground and may pose a potential 
electrostatic charging hazard: (Metal probe - 19.5pF)   

                     

In the event that any part of the instrument should become electrostatically charged in a potentially 
hazardous location, follow company policies for testing and clearing the area of any hazardous gases before 
attempting to use the instrument in that location. If this is not possible allow sufficient time for the instrument to 
naturally dissipate any charges. Given the atmosphere, this could take several hours. 

 

 

SAFETY APPROVALS FOR TL3-A, TL3-W AND TL3-R:   

 
 
 
 
 

 

 
Ex ia IIC T4 Ga

 

 

 

Applicable Standards are:    

 

 

Agency or Safety Designation 

 

IEC 60079-0:Ed 7, IEC 60079-11:Ed 6      

 

IECEx: Internationally Recognized Body 

 
EN 60079-0:2018,  EN60079-11:2012  

 

ATEX: Europe 

 

 
 
 
 

www.

.com

[email protected]

1.800.561.8187

Содержание ThermoProbe TL3-A

Страница 1: ...11 2020 JK ThermoProbe Models TL3A TL3W and TL3R TL3 A TL3 W TL3 R USER MANUAL www com information itm com 1 800 561 8187...

Страница 2: ...ied Body Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul Naviglio MI Italy Identification Number 2575 Auditing body QAN Intertek Testing Certification Limited Intertek House Cleeve Road...

Страница 3: ...tery is aligned with symbol embossed in the battery case Batteries must not be installed with polarity reversed where one cell could charge another cell a Ensure the instrument is in a non hazardous a...

Страница 4: ...r the following selections C F Select Celsius or Fahrenheit temperature display DEC Select 0 1 or 0 01 display resolution 0 01 0 001 for TL3 R LAH Displays Lowest reading LO Average reading AVG and Hi...

Страница 5: ...be performed by qualified personnel with proper reference instrumentation Zero Correction is intended to be performed at water freezing point 0 C 32 F The Zero Correction can be accessed by selecting...

Страница 6: ...ot be used until repairs have been made c The instrument should be clean for safety and ease of use d The physical measurement location should be evaluated for primary and secondary risks e Power sour...

Страница 7: ...An intrinsically safe system assumes the fuel and oxygen are present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument loop can never b...

Отзывы: