Lowest, Highest & Average readings 

The lowest, highest and average temperatures recorded while the TL3 is powered on can be accessed from the menu. 
These are indicated by LO, AVG or HI indicators shown on the display immediately before the corresponding reading 
is displayed. These readings are deleted after the unit is powered off. 
 

Bluetooth

®

 Wireless (TL3-W or TL3-R models only)

 

Bluetooth can be enabled from the menu to transmit temperature data to supported Bluetooth devices.  Please go to 

Display Codes  

OPEN Ct

 - Indicates the sensor is operating above its temperature limit, or the sensor is damaged.

 

Shrt Ct

 - Indicates the sensor is operating below its temperature limit or the sensor is damaged. 

  

Lo bAtt

 - Indicates the battery voltage is low. 

 

 

 
 
 

Zero Correction (TL3-R model only) 
 

Note: A password (‘247’) is required to enter Zero Correction mode. 

Corrections made will offset temperature over the entire range of the device. 
Zero correction should only be performed by qualified personnel with proper  
reference instrumentation. 
Zero Correction is intended to be performed at water freezing point 0°C (32°F). 

 

The Zero Correction can be accessed by selecting ZERO from the menu. 
From the Zero Correction menu, selecting Corr will display the current zero correction.   
 

 
 
 
 
 
 
 

 
 
Selecting AdJ will display a password field of 3 digits. Enter the password (247) one digit at a time using the 
ADVANCE button to increment each display digit. Then use the SELECT button to go to the next digit. Pressing 
SELECT for the 3

rd

 digit will take you to a temperature display where you can adjust the displayed temperature. The 

arrow indicator shows the direction you are changing the temperature. Press and hold the ADVANCE button to 
increment the value by 0.001 degrees until it reaches 10 and then it will increment at a rate of 0.01 degree per 
second. Use the SELECT button to go from adjust UP, adjust DOWN, Save correction, and Cancel (don’t save). Once 
you have saved the adjustment you can view the zero adjustment being applied from the Zero→Corr→’Displays 
current correction’ menu selection. 

 

 

 

www.

.com

[email protected]

1.800.561.8187

Содержание ThermoProbe TL3-A

Страница 1: ...11 2020 JK ThermoProbe Models TL3A TL3W and TL3R TL3 A TL3 W TL3 R USER MANUAL www com information itm com 1 800 561 8187...

Страница 2: ...ied Body Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul Naviglio MI Italy Identification Number 2575 Auditing body QAN Intertek Testing Certification Limited Intertek House Cleeve Road...

Страница 3: ...tery is aligned with symbol embossed in the battery case Batteries must not be installed with polarity reversed where one cell could charge another cell a Ensure the instrument is in a non hazardous a...

Страница 4: ...r the following selections C F Select Celsius or Fahrenheit temperature display DEC Select 0 1 or 0 01 display resolution 0 01 0 001 for TL3 R LAH Displays Lowest reading LO Average reading AVG and Hi...

Страница 5: ...be performed by qualified personnel with proper reference instrumentation Zero Correction is intended to be performed at water freezing point 0 C 32 F The Zero Correction can be accessed by selecting...

Страница 6: ...ot be used until repairs have been made c The instrument should be clean for safety and ease of use d The physical measurement location should be evaluated for primary and secondary risks e Power sour...

Страница 7: ...An intrinsically safe system assumes the fuel and oxygen are present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument loop can never b...

Отзывы: