 USER INSTRUCTIONS – TL3-A, TL3-W & TL3-R 

     

 

INTRODUCTION  

 

This manual describes the basic function, use and safety instructions for the model TL3-A, TL3-W and TL3-R portable 
digital reference thermometer instruments. 

 

  

REPLACING BATTERY  

When the battery voltage is low, the display will show ‘Lo bAtt’.  
 
Replace the battery as soon as possible in a safe location after the low battery is noticed. This will avoid possible 
malfunctioning. Do not attempt to calibrate the instrument if the low battery indicator is displayed.  

 

WARNING:  

 

 

Batteries must be changed in Non-hazardous area.   

 

Batteries must be of correct approved type.   

 

Batteries must be installed with correct polarity making sure the (+) end of the battery is aligned with (+) 
symbol embossed in the battery case.  

 

Batteries must not be installed with polarity reversed where one cell could charge another cell.   
 

a) Ensure the instrument is in a non-hazardous area & powered off. 
b) Use a #2 Phillips drive to remove the 1 screw holding the back cover on the TL3. 
c) Replace each new battery making sure the (+) end of the battery is aligned with (+) symbol embossed in the 
battery case. 
d)

 

Reinstall the cover. Make sure the rubber gasket is seated in its groove around the case opening (TL3-W, TL3-R). 

  

CERTIFIED

 Batteries for the TL3-A, TL3-W and TL3-R are as follows:  

 

Model    

 

Type 

 

 

 

Duracell 

 

Panasonic 

 

TL3-A    

 

(1) CR2032 LiMnO

2

 

 

DL2032  

 

CR2032 

 

 

 

TL3-W, TL3-R   

(2) AAA (LR03) Alkaline  

MN2400 

 

LR03XWA 

 

AUTHORIZED REPAIR 

 

It is recommended that service beyond the scope of this manual be performed by ThermoProbe, Inc. or one of its 
authorized distributors.  

 
 
 
 

 

 
 
 
 
 

www.

.com

[email protected]

1.800.561.8187

Содержание ThermoProbe TL3-A

Страница 1: ...11 2020 JK ThermoProbe Models TL3A TL3W and TL3R TL3 A TL3 W TL3 R USER MANUAL www com information itm com 1 800 561 8187...

Страница 2: ...ied Body Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul Naviglio MI Italy Identification Number 2575 Auditing body QAN Intertek Testing Certification Limited Intertek House Cleeve Road...

Страница 3: ...tery is aligned with symbol embossed in the battery case Batteries must not be installed with polarity reversed where one cell could charge another cell a Ensure the instrument is in a non hazardous a...

Страница 4: ...r the following selections C F Select Celsius or Fahrenheit temperature display DEC Select 0 1 or 0 01 display resolution 0 01 0 001 for TL3 R LAH Displays Lowest reading LO Average reading AVG and Hi...

Страница 5: ...be performed by qualified personnel with proper reference instrumentation Zero Correction is intended to be performed at water freezing point 0 C 32 F The Zero Correction can be accessed by selecting...

Страница 6: ...ot be used until repairs have been made c The instrument should be clean for safety and ease of use d The physical measurement location should be evaluated for primary and secondary risks e Power sour...

Страница 7: ...An intrinsically safe system assumes the fuel and oxygen are present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument loop can never b...

Отзывы: