If the starting type was set to 

PC Start

, then the 

Start

 key will 

become active. Click to 

Start

.

If the starting type was set to 

Button Start

, then the logger 

display will indicate that the logger button has to be pressed 

to start logging. Keep the button pressed for about 3 seconds.

3 sec.

If the starting type was set to 

Date/Time Start

, then the 

logger will automatically start logging at the programmed time.

Button Functions

Pressing the logger button will give different outputs depending 

on the logger status.
• If the logger is in “

WAIT TO LOG

” status, then pressing 

the button will navigate through current measurement, high 

and low alarm thresholds.

• If the logger is in “

LOG

” or “

LOG DONE

” status, then 

pressing the button will navigate through current measure-

ment, maximum and minimum recorded values.

• If you keep the button pressed, after 2 seconds the display 

will show alternatively (toggle every 2 seconds) the high 

and low alarm thresholds.

keep pressed,

then after 2 sec.

toggle every

2 seconds

Notes:

 If an alarm threshold has not been 

set, then the logger will display the 

diS

” (disabled) message.

 If an alarm condition has occurred during logging, then 

in the “

LOG

” or “

LOG DONE

” status the corresponding 

alarm indication (

“ALARM HI

” or “

ALARM LO

”) will blink.

Operational Guide

Technical Service

This instrument is factory calibrated, but it is generally 

recommended to have all thermometers recalibrated at least 

once a year. 

Initial Operation

The logger is supplied complete with a CR2032 lithium battery. 

Open the battery compartment cover (see “Battery Replacement” 

section for details), insert the battery with the + sign facing 

up and tighten back the compartment cover.
The display will show the 

“PC”

 message 

to indicate that the logger needs to be 

programmed through a PC.

Startup

In order to communicate with the logger, the application 

software must be installed on your PC. 
Connect 

HI144002

 communication cradle to your computer 

through a USB port, and place the 

HI144

 on the cradle.

Run the application software. The yellow LED on the cradle 

blinks to indicate that the communication is active. 

Do not 

remove the logger while the LED is blinking

The following operating parameters can be set:
• Measurement Unit: °C or °F

• Starting Type: PC Start, Button Start or Date/Time Start 

(hh:mm:ss)

• Logging Method: Wrap Around Memory, until memory is full (at 

8000 samples) or number of samples (a maximum of 8000)

• Logging Interval: from 1 minute to 24 hours (1 minute 

resolution)

• High and Low Alarm thresholds

• Logger Password
Click on the 

Program

 key to program the logger with the 

selected options.