Configuration Commands
Copyright© Itech Electronic
Co., Ltd.
19
Default Value
ON
Example
RES:RANG 0
Query Syntax
RESistance:RANGe[:STATe]?
Return Parameters
0|1
Relevant Command
RES:RANG?
[CHANnel:]FREQuency[:STATe]
This command selects the excitation frequency for measurement. 0 indicates
500HZ, and 1 indicates 1000HZ.
Command syntax
[CHANnel:]FREQuency[:STATe] <bool>
Parameters
0 | 1 or ON| OFF
Default Value
ON
Example
FREQuency 0
Query Syntax
FREQuency [:STATe]?
Return Parameters
0 | 1
Relevant Command
FREQuency?
[CHANnel:]PANEl[:STATe]
This command switches the measurement mode to front panel or rear panel. 1
indicates rear panel. 0 indicates front panel.
Command syntax
[CHANnel:]PANEl[:STATe] <bool>