![ITech IT5101 Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/itech/it5101/it5101_programming-manual_2095395029.webp)
Configuration Commands
Copyright© Itech Electronic
Co., Ltd.
21
When the compare function is turned on, if you turn on the autorange, the compare
function will turn off automatically.
Command syntax
AUTorange <1|0|ON|OFF>
Query syntax
AUTorange?
Arguments
1|0|ON|OFF
Example
AUTorange ON
Query command
AUTorange?
Arguments
ON
AUTorange:VOLTage <1|0|ON|OFF>
This command is used to set and query the autorange of voltage. When
the compare function is turned on, if you turn on the autorange, the
compare function will turn off automatically.
Command syntax
AUTorange:VOLTage <1|0|ON|OFF>
Query syntax
AUTorange:VOLTage?
Arguments
1|0|ON|OFF
AUTorange:RESistance <1|0|ON|OFF>
This command is used to set and query the autorange of resistance.
When the compare function is turned on, if you turn on the autorange,
the compare function will turn off automatically.
Command syntax
AUTorange:RESistance <1|0|ON|OFF>
Query syntax
AUTorange:RESistance?