Configuration Commands
Copyright© Itech Electronic
Co., Ltd.
19
FUNCtion <RV|RESistance|VOLTage>
This command is used to set and query measurement function.
Command syntax
FUNCtion <RV|RESistance|VOLTage>
Query syntax
FUNCtion?
Arguments
RV|RESistance|VOLTage
Example
FUNCtion RESistance
Query command
FUNCtion?
Arguments
RESistance
RESistance:RANGe <0-3100>
RESistance:RANGe<0-3.1>
This command is used to set and query the range of resistance.
The
resistance range of IT5101/IT5101H model and IT5101E model is not the same.
The command of IT5101/IT5101H is:
RESistance:RANGe <0-3100>, The
command of IT5101E is: RESistance:RANGe<0-3.1>
。
Command syntax
RESistance:RANGe <0-3100>
RESistance:RANGe<0-3.1>
Query syntax
RESistance:RANGe?
Arguments
IT5101/IT5101H:
3.0000E-3/30.000E-3/300.00E-3/3.0000E+0/30.000E+0/300.00E+0/3.0
000E+3
IT5101E:300.00E-3/3.0000E+0
Example
RESistance:RANGe 120E-3