Load Commands
Copyright © Itech Electronic Co., Ltd.
46
Default value
MAXimum
Example
RES:TRIG MAX
Query syntax
[SOURce:]RESistance[:LEVel]:TRIGgered[:AMPLitude]? [MINimum|MAXimum]
Returns
NR3
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
<NRf+>
This command sets the input voltage value under CV mode.
Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <NRf+>
Arguments
<MINimum-MAXimum>|MINimum|MAXimum
Default value
MINimum
Example
VOLT 10.0
Query syntax
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MINimum|MAXimum]
Returns
NR3
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
<NRf+>
This command sets the input voltage value when the electronic load receives a
trigger signal.
Syntax
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <NRf+>