Under PE+ 2.0 mode, the voltage range is 5V-20V. The user can click
Set button to increase or decrease the voltage value with an adjustment
step of 0.5V. Or, the user can press and hold Set button for fast adjust-
ment or directly input a voltage value.
Under PE+ 2.0 mode, the user also needs to set the current duration.
This function is the same as that under the PE+ mode.
•
Under USB PD mode, the software will distinguish USB PD2.0 or USB
PD3.0 based on the protocol supported by the DUT. The user can define
the
Position
value to select the voltage and current position.
The USB PD3.0 has PPS specifications. When defining the
Position
val-
ue and selecting
PPS
mode, the user can customize the voltage and cur-
rent value. This function can meet more test needs of the user.
5. Set the load mode and corresponding parameter value. For load mode and
related description, please refer to corresponding user manual.
When the test mode is PE+ or PE+ 2.0, the loading current of the load
cannot be less than 0.3A. Otherwise, the voltage setting will automatically
recover to 5V.
6. Click
START
, and change the output voltage based on setting.
7. Click
ON/OFF
to start load input control. When the load input state light turns
green and displays ON, start loading test.
When the existing mode is PE+ or PE+ 2.0, press
START
to change the
output voltage, and then turn on the load input control.
8. Observe the
Readback
value at the right side. If you want to save data,
check
Record
to record test data.
2.3.2.3 Test under List Mode
The user can edit 1~99 steps of list file in the software for automatic execution
of continuous tests.