background image

7

MENU FUNCTIONS - Continued

to save. Press the CANCEL button to abandon changes. While viewing pressure / temperature 

charts, use the UP / DOWN ARROW buttons to scroll through the chart. 

Hide / show elapsed time: In Normal Mode, the elapsed time is shown in the lower-left corner 

of the display. The elapsed time may be turned off by selecting this menu item and Pressing the 

ENTER button. If disabled, the menu item will be shown as “Show Elapsed Time”If enabled, 

the menu item will be shown as “Hide Elapsed Time”.

Disable / enable auto shut-off: When Auto Shut-off is enabled, the MFM300 will power-off 

in 15 minutes if no buttons have been pressed. To disable Auto Shut-off, use the UP / DOWN 

ARROW buttons to highlight this menu item and press the ENTER button. Auto Shut-off is the 

default setting when turning the MFM300 on.

Setup:  Press the MENU button, highlight the “Setup” option and press the ENTER button. 

Setup Items include set altitude, set units of measure, temperature calibration, and zero-out 

pressure transducers. Use the UP / DOWN ARROW buttons and press ENTER to select the 

highlighted option. To return to the main menu, press CANCEL at any time.

•   Set altitude: Use the UP / DOWN ARROW buttons to enter the current altitude in 500-foot  

 increments.

•   Units of measure: Select either English or Metric units of measure.

•   Temperature calibration: Place a temperature probe connected to T1 into an  

 

 environment with a known temperature and adjust the reading to calibrate and match.

•   Zero-out probe P1, Zero-out probe P2: When a pressure transducer is attached to U1 or  

 U2, select this option before connecting to the system if the transducer reading is not “0  

 PSI” after setting the altitude. 

USING THE MFM300 WITH A GL DATA LOGGER

Connect the Data Logger Reader to U1 using the GL300 data logger cable. 

NOTE: THE GL300 DATA LOGGER CABLE IS NOT SUPPORTED ON U2. 

 

Attach the GL100 Data Logger Tag to the reader. The MFM300 detects the Data Logger and 

displays the GL100 Data Log Menu. If the Data Logger has been previously programmed for 

a mission, the mission description is displayed. Below the mission description are the menu 

options. Use the UP / DOWN ARROW buttons to scroll to the desired menu option and press the 

ENTER button.

Check settings: Select this menu option to view the current mission status.

The status screen displays the following data:

  •  Mission description or “Name”

  •  Sampling status: Active or stopped

  •  Sample interval: Time between samples

  •  Mission start time and date

  •  Action when Data Logger is full: Stop or rollover

  •  Record count

Press any key to return to the GL100 Data Log Menu.

 

Program a mission: Enter up to 20 alphanumeric characters to describe the pending mission. 

The symbol in the lower-left corner of the display indicates whether letters or numbers are 

entered. “ABC” indicates upper case letters, “abc” indicates lower case letters and “1 2 3” 

indicates numbers.

Содержание Cooper MFM300

Страница 1: ...MFM300 Multi Function Instrument User Guide and Instrument Warranty...

Страница 2: ...te Superheat Sub Cooling 5 Measure Dynamic System Superheat Sub Cooling 5 6 Measure Relative Humidity and Dry Bulb Temperature 6 Display Psychometrics 6 Measure Air Flow 6 Air Balancing 6 Navigate the...

Страница 3: ...JACK U1 UNIVERSAL JACK U2 POWER TEMPERATURE JACK T2 TEMPERATURE JACK T1 DOWN ARROW PSYCHROMETRICS PSYCH TEMPERATURE JACK T3 SHIFT SUPERHEAT SUB COOLING SH UP ARROW ENTER REFRIGERANT REF MENU AVERAGE...

Страница 4: ...ssage will appear on the display DISPLAY MIN MAX AND AVG VALUES When in Normal Mode press MIN button to display the lowest readings sensed by each probe MAX button to display the highest readings sens...

Страница 5: ...memory There is room in the memory for up to five more refrigerants New refrigerant P T data can be downloaded from the Cooper Atkins website www cooper atkins com mfm asp using the MFM300 Toolbox an...

Страница 6: ...layed you must enter the shape and size of the duct opening where airflow is being measured To do this press the MENU button while viewing CFM Use the on screen directions to enter the duct shape and...

Страница 7: ...Place a temperature probe connected to T1 into an environment with a known temperature and adjust the reading to calibrate and match Zero out probe P1 Zero out probe P2 When a pressure transducer is...

Страница 8: ...ace it where temperatures are to be measured Download data Select this menu option to store the Data Logger contents in the MFM300 and view at a later time View data Select this menu option to view th...

Страница 9: ...rature Probes Temperature Measurement Range 58 to 302 F 50 to 150 C Temperature Accuracy 0 3F 0 2C or 0 5 of Reading whichever is greater Temperature Display Resolution 0 1 Degree Relative Humidity RH...

Страница 10: ...es notice that any statements or representations made by it will not create a warranty that this product is fit for any particular purpose Such statements representations or descriptions are informati...

Отзывы: