background image

5

MEASURE TEMPERATURE DIFFERENCE (Delta T) Continued 

  FIRST PRESS: 

T1 – T2;

  SECOND PRESS:  T1 – T3;

  THIRD PRESS:  T2 – T3;

  FOURTH PRESS:  DELTA T DISPLAY OFF.

MEASURE PRESSURE

Connect a Cooper Pressure Transducer to either U1 or U2 (or both). The MFM300 will sense 

the probe(s) and display the pressure measurement(s) with the appropriate U1 or U2 label.  

Before connecting the transducer(s) to the cooling system, zero the pressure reading(s) (refer 

to Setup section on zeroing the pressure probes and setting altitude). 

NOTE: AT SEA LEVEL, THE PRESSURE READING SHOULD BE 0 PSI (0 KPA) WHEN NOT   

 

CONNECTED TO A COOLING SYSTEM. AT ELEVATIONS ABOVE SEA LEVEL THE    

 

DISPLAY WILL SHOW A NEGATIVE PRESSURE READING (-0.5 PSI / 1000  

 

 

FEET ALTITUDE). CONNECT THE TRANSDUCER(S) TO THE COOLING SYSTEM ACCESS  

 

PORTS, USING BEST INDUSTRY PRACTICES, TO BEGIN MEASURING SYSTEM    

 PRESSURE.

MEASURE SYSTEM SUPERHEAT / SUB-COOLING

 

Select the refrigerant: The selected refrigerant is displayed when in Superheat / Sub-cooling 

Mode, or when viewing pressure / temperature charts. To change the refrigerant, press the REF 

button. The “Up / Down Arrows” will appear beside the refrigerant name. Use the UP / DOWN 

ARROW buttons to scroll to the desired refrigerant and then press the ENTER or REF button. 

Press the CANCEL button to abandon changes.

 

Add and / or change refrigerants: The MFM300 stores P / T data from 15 popular refrigerants 

that reside in permanent memory. There is room in the memory for up to five more 

refrigerants. New refrigerant P / T data can be downloaded from the Cooper-Atkins website 

(www.cooper-atkins.com/mfm.asp) using the MFM300 Toolbox and the optional computer 

connection cable. 

CALCULATE SUPERHEAT / SUB-COOLING

The MFM300 can be used as a Superheat / Sub-cooling calculator. Connect the temperature 

probes to the system and press the SH button. Select the refrigerant and press the ENTER 

button. Since there are no pressure transducers connected, system pressure will default to  

25 PSI for superheat and 75 PSI for Sub-cooling. You may manually enter the actual pressure 

value from the manifold gauge into the MFM300 using the UP / DOWN ARROW buttons. Once 

the pressure is entered for Superheat, use the SHIFT button to move to Sub-cooling, where 

pressure is manually entered the same way. Remember that the values shown in this mode are 

not dynamic.

MEASURE DYNAMIC SYSTEM SUPERHEAT / SUB-COOLING

Measuring dynamic system Superheat or Sub-cooling requires both a temperature probe 

and a pressure transducer. The refrigerant type selected on the display must match the 

refrigerant type used in the system. Refer to the section Select the Refrigerant. For Superheat 

measurement, connect the temperature probe to T1, and the pressure transducer to U1. For 

Sub-cooling measurement, connect the temperature probe to T2 and the pressure transducer 

to U2. To measure superheat, attach the temperature probe to the system suction line near the 

compressor. Attach the pressure transducer to the low side access port. Press the SH button. 

Содержание Cooper MFM300

Страница 1: ...MFM300 Multi Function Instrument User Guide and Instrument Warranty...

Страница 2: ...te Superheat Sub Cooling 5 Measure Dynamic System Superheat Sub Cooling 5 6 Measure Relative Humidity and Dry Bulb Temperature 6 Display Psychometrics 6 Measure Air Flow 6 Air Balancing 6 Navigate the...

Страница 3: ...JACK U1 UNIVERSAL JACK U2 POWER TEMPERATURE JACK T2 TEMPERATURE JACK T1 DOWN ARROW PSYCHROMETRICS PSYCH TEMPERATURE JACK T3 SHIFT SUPERHEAT SUB COOLING SH UP ARROW ENTER REFRIGERANT REF MENU AVERAGE...

Страница 4: ...ssage will appear on the display DISPLAY MIN MAX AND AVG VALUES When in Normal Mode press MIN button to display the lowest readings sensed by each probe MAX button to display the highest readings sens...

Страница 5: ...memory There is room in the memory for up to five more refrigerants New refrigerant P T data can be downloaded from the Cooper Atkins website www cooper atkins com mfm asp using the MFM300 Toolbox an...

Страница 6: ...layed you must enter the shape and size of the duct opening where airflow is being measured To do this press the MENU button while viewing CFM Use the on screen directions to enter the duct shape and...

Страница 7: ...Place a temperature probe connected to T1 into an environment with a known temperature and adjust the reading to calibrate and match Zero out probe P1 Zero out probe P2 When a pressure transducer is...

Страница 8: ...ace it where temperatures are to be measured Download data Select this menu option to store the Data Logger contents in the MFM300 and view at a later time View data Select this menu option to view th...

Страница 9: ...rature Probes Temperature Measurement Range 58 to 302 F 50 to 150 C Temperature Accuracy 0 3F 0 2C or 0 5 of Reading whichever is greater Temperature Display Resolution 0 1 Degree Relative Humidity RH...

Страница 10: ...es notice that any statements or representations made by it will not create a warranty that this product is fit for any particular purpose Such statements representations or descriptions are informati...

Отзывы: