Fig 5.3 Trigger Mode / Slave
5.2.3
Overlap mode
Camera keeps standby status until trigger signal is entered and Readout starts after
exposure process set earlier if trigger input occurs. When new trigger input occurs
during Readout of First image, it keeps Readout and pass exposure process of new
trigger input. Provided, however, that when trigger input occurs during Exposure since
Exposure Time is longer than trigger interval, that trigger signal is ignored. To obtain
the image as maximum frame for trigger input, Exposure Time should not be longer
than Readout time, trigger time should not be shorter than Readout time. Readout time
for the IC-M29 is 224.7ms.(Quad Output) Overlap mode operates ideally when trigger
signal interval or exposure time is constantly kept. Overlap mode only operating in the
Trigger mode.
Fig 5.4 Overlap Mode Operation(Transfer Time > Exposure Time)