background image

 

                                                                                                                                                      

26   

To check the Address value press the scroll key until the lower display indicates, 

ADDR 
 

The upper display will show the current value that can be modified with the UP and DOWN keys. 

 

14.8

 

Monitoring The Indicator Status 

For the indicator the REM beacon is lit continuously when the user calibration is active, the REM 

beacon flashes when the PC communications port is active. 

 

14.8.1

 

Units 

 

Momentary pressing of the Scroll key will show the controller units °C or °F. 

 

14.9

 

Indicator Operation 

The indicator can be configured with up to five custom calibration points; the points contain "data 

pairs". First the temperature (point) and secondly the Error (offset) at this temperature point. 

Isotech Dry Block calibration certificates will show the values to suit a particular sensor. 

 

These values can be inspected, and modified with the following procedure, 

 

Hold the PAGE key down until the display indicates, 

CONF  

GOTO 

 

then 

 

CODE 

 

Set the Code to 2 with the UP key 

 

The display reads, 

CONF 

 

Press the Page Key until the controller shows 

CAL 

 

Now use the Scroll key to examine the data pairs. The values Pnt 1- Pnt 5 and Ofs 1 to Ofs 5 

can be modified with the UP and DOWN keys.  
 

To exit this mode hold the Page key until the top display shows, 

CONF 

GOTO 

 

And  then  set  the  upper  display  to  Lev  1.  While  in  this  mode  take  care  not  to  modify  other 

parameters. 

 

 

 

Содержание VENUS 4951 BASIC

Страница 1: ...ited Pine Grove Southport Merseyside PR9 9AG England Telephone 44 0 1704 543830 Fax 44 0 1704 544799 Email info isotech co uk Website www isotech co uk Venus 4000 Series 02 04 17 User Maintenance Manu...

Страница 2: ...IS GUARANTEE INTERFERENCE WITH OR FAILURE TO PROPERLY MAINTAIN THIS INSTRUMENT MAY INVALIDATE THIS GUARANTEE The company is always willing to give technical advice and assistance where appropriate Equ...

Страница 3: ...C OPERATION 14 10 3 SITE OR SELF CONTAINED CALIBRATORS 14 10 4 EXTERNAL STANDARDS BASIC 15 10 5 ADVANCED CALIBRATORS 15 11 MODE OF OPERATION 16 11 1 METAL BLOCK BATH 16 11 2 BLACKBODY SOURCE 16 11 3 S...

Страница 4: ...8 1 Units 26 14 9 INDICATOR OPERATION 26 15 CALIBRATION DATA EXAMPLE 27 15 1 CONNECTING A 4 20MA CURRENT TRANSMITTER 27 15 2 USING THE CURRENT LOOP INTERFACE 935 06 161 WITH THE SITE MODEL 29 15 3 TES...

Страница 5: ...u are advised after unpacking the unit to inspect it for any sign of shipping damage and confirm that your delivery is in accordance with the packing check list If you find any damage or that part of...

Страница 6: ...se yourself with all warnings hazards and safety information Regularly inspect the equipment accessories and electrical leads Do not use if there is any sign of damage Keep the equipment clean only us...

Страница 7: ...ual before use IEC 417 Caution hot surface risk of burn O Off l On Waste Electrical and Electronic Equipment WEEE directive symbol Complies with European regulations CE mark USB symbol Ethernet RJ45 s...

Страница 8: ...lease familiarise yourself with the parts of the manual relevant to your model Your unit s supply voltage requirement is specified on a label on the instrument along with the serial number Only connec...

Страница 9: ...h a protective earth If using an alternative cord ensure it is of adequate rating If fitting a plug to the cable we have supplied the cable is colour coded as follows Colour Function Green yellow Eart...

Страница 10: ...ic Compatibility EMC 89 336 EEC as amended by EC Directive 92 31 EEC and the European Low Voltage Directive 73 25 EEC amended by 93 68 EEC To ensure emission compliance please ensure that connecting l...

Страница 11: ...rained about the selection and use of liquids over a wide temperature range It is expected that personnel using this equipment will be competent with the management of apparatus which may be powered o...

Страница 12: ...ee space above the unit to avoid risk of burn or fire The equipment should only be used indoors Site in a way that allows access to the power switch and to allow the cable to be disconnected Avoid exc...

Страница 13: ...13 9 Consider The Environment The equipment should be recycled or disposed of in a proper way following the Waste Electrical and Electronic Equipment WEEE directive...

Страница 14: ...he block when the set temperature is lower than the block temperature 10 1 Comparison Calibration By definition one compares industrial thermometers to a calibrated standard There are three methods co...

Страница 15: ...e indicator may be used with a range of blocks This method also meets ISO9000 requirements 10 5 ADVANCED Calibrators The ADVANCED model includes an indicator with channels for both the test thermomete...

Страница 16: ...average values accordingly 11 2 Blackbody Source The Blackbody Target Kit 951 06 04 can be added for simple calibration on infrared thermometers Place the target into the well with the supplied therm...

Страница 17: ...port Kit part number 921 02 06 allows thermometers with a diameter of 2 to 6 5mm to be suspended in the liquid b The Sensor Guide which fits into the tank and can support a number of thermometers part...

Страница 18: ...ch Once the Power lead is connected this switch will power up and down the equipment 3 Fuse carrier and window showing fuse rating With the mains cable disconnected the fuse carrier can be removed for...

Страница 19: ...emoved for the inspection or replacement of the fuse The fuse rating is displayed through the window 4 PRT input socket These allow for the connection of external Platinum resistance thermometers Warn...

Страница 20: ...l Liquid Bath Mode 0 02 C Uniformity Lower 40mm Axial As Liquid Bath Mode 0 026 C Cools from 140 C to 0 C in 15 minutes Heats from 30 C to 140 C in 15 minutes Best Performance See Graph refer to Secti...

Страница 21: ...raphs 13 1 Heat Up Cool Down 13 2 Sigma Uncertainties Best uncertainty using Basic model with a TTI 2 and a 935 14 82 calibrated system Site version uncertainty Site model using built in indicator at...

Страница 22: ...e it is advantageous to limit the heating or cooling rate An example might be when testing bimetallic thermostats by forcing the Dry Block to heat at a controlled rate it is easier to determine the te...

Страница 23: ...e that can be modified with the UP and DOWN keys 14 3 Monitoring The Controller Status A row of beacons indicate the controllers status as follows OP1 Heat Output OP2 Cool Output only for models which...

Страница 24: ...ith three wire PRT connections ensure identical lead resistance in all three legs of the connection by using the same length and gauge of cables 14 5 3 Connecting Thermocouples Set the input type to s...

Страница 25: ...tion can make a significant difference to the accuracy of the instrument and this REM beacon provides a clear and continuous indication of the calibration status Isotech will configure and set user ca...

Страница 26: ...e points contain data pairs First the temperature point and secondly the Error offset at this temperature point Isotech Dry Block calibration certificates will show the values to suit a particular sen...

Страница 27: ...The transmitter should be powered externally a 2 49 Ohm current sense resistor is fitted internally to the equipment and this allows the indicator to read mA input To scale the linear to display temp...

Страница 28: ...28...

Страница 29: ...Connect the Transmitter to the equipment as shown in the above image Note do not connect 935 06 161 via the PRT socket The Current Loop interface has built in resistance and does not require the inter...

Страница 30: ...eversal of the polarity for the Switch Contact feature 1 Enter Configuration level on the temperature Indicator as described in INDICATOR OPERATION 2 Press the PAGE button to enter IO 1 3 Press the Sc...

Страница 31: ...calibration baths may be connected together in a daisy chain configuration and then linked to a single RS232 see diagram Note The RS 422 standard specifies a maximum lead length of 1200M 4000ft A tru...

Страница 32: ...the milliK or TTI 10 The software requires Windows 9X XP a minimum of 5Mb of free hard drive space and free serial ports for the instruments to be connected 17 1 Development Cal Notepad was developed...

Страница 33: ...ally if a license is purchased it allows for fully automatic calibration including certificate printing and the calculation of coefficients Refer to the Isotech I Cal EASY handbook and help system for...

Страница 34: ...revious configuration E CaL Calibration error Re instate Factory calibration refer to Isotech E2 Er EEPROM error Return to Isotech for repair EE Er Non vol memory error Note the error and contact Isot...

Страница 35: ...nGL Process Value Low Limit 0 00 Config IP Name Description Value inPt Linearisation Type RTD CJC CJC Type EXT imP Sensor break impedance AUTO 1 Config CAL Name Description Value UCAL User Calibration...

Страница 36: ...Ltch4 Alarm 4 Latching NO 0 Config HA Name Description Value id Module Identity CMS 7 Func Module Function CMS 65 bAud Baud Rate 9600 0 Prty Comms Parity NONE 0 rES Comms Resolution FUL 0 Config 1A N...

Страница 37: ...eter Support Kit 921 02 06 Gantry 921 02 07 Blank Disc 951 02 20A Drilled Disc 951 02 20 3 x Gantry Plugs 820 20 24 3 x O Seals 932 12 49 Option 4 Blackbody target 951 06 04 Option 5 Surface Sensor In...

Отзывы: