Page 9 of 12 

TTI-10 Iss.02 – 07/11 

Channel Activation / [Chnl] 

 

To  activate  or  deactivate  a  measuring  channel  push  [ENTER/MENU].  Use  the  up  and  down  keys  [



]  to  select 

Chnl

. Push again [ENTER/MENU]. On the left corner of the display appears a small 

1

, which indicates the selected 

channel. To change the channel use the up and down keys [



]. Push [ENTER/MENU] to confirm. Use the up and 

down keys [



] again to activate 

on

 or deactivate 

off

 the requested measuring channel and push [ENTER/MENU] to 

confirm. Push [ESC] to revert back to the measuring mode. 

 

Note:

 As a minimum one channel is active! 

 

 

MENU 

 

 

 

 

 

 

Unit    Lin2   

CAL

*

 

  Chnl    Log   

*

The calibration is hidden when the instrument has been calibrated at Isotech, see page 6 for details.

 

Memory Setup/Lo6 

 

The TTI-10 can log up to 4000 measurements in memory and report the maximum, minimum and average values of 

the logged data. 

 

Push [ENTER/MENU] and use the up and down keys [



] to select 

Lo6

. Push [ENTER/MENU] to confirm. Use the 

up and down keys [



] again to start 

[on]

 or Stop 

[off]

 the logger. Push [ENTER/MENU] to confirm. Use the up 

and down keys [



] again to select between automatic storage 

[Auto]

 or manually operated storage 

[SPot]

. Push 

[ENTER/MENU] to confirm]. Use the up and down keys [



] again to select between to add on data 

[Add] 

and 

creating a new file

[nLo6]

. Push [ENTER/MENU] to confirm]. By selected automatic storage at the end you have to 

select the time interval: 

 

1 S    1 second 

5 S    5 seconds 

10 S   10 seconds 

20 S   20 seconds 

30 S   30 seconds 

1 M   1 minute 
2 M   2 minutes 

5 M   5 minutes 

10 M  10 minutes 

20 M  20 minutes 

 

By selected manually operated storage you are able to save the measurement by pressing ESC by each time. 

 

Push [ENTER/MENU] to confirm. Push [ESC] to revert back to the measuring mode. 

 

 

MENU 

 

 

 

 

 

 

Unit    Lin2   

CAL

*

 

  Chnl    Log   

*

The calibration is hidden when the instrument has been calibrated at Isotech, see page 6 for details.

 

E.g. of the instrument’s LCD by activated logging mode: 

 

On the bottom you can see the percentage of the occupied memory (0..99%). If a 

calibration option is activated the display is alternating between displaying memory status 

and calibration information. 

Содержание TTI-10

Страница 1: ...4 543830 Fax 44 0 1704 544799 Internet www isotech co uk E mail info isotech co uk The company is always willing to give technical advice and assistance where appropriate Equally because of the progra...

Страница 2: ...C F and Ohms 6 Temperature Difference Mode Line2 6 Calibration function CAL 6 Channel Activation Chnl 9 Memory Setup Lo6 9 Recalling the memory data HOLD MAX MIN AVE 10 Measuring rate Normal Mode FAS...

Страница 3: ...PARTS ARE NOT COVERED BY THIS GUARANTEE INTERFERENCE WITH OR FAILURE TO PROPERLY MAINTAIN THIS INSTRUMENT MAY INVALIDATE THIS GUARANTEE RECOMMENDATION The life of your ISOTECH Instrument will be prol...

Страница 4: ...ERSONNEL FAMILIAR WITH GOOD MEASUREMENT PRACTICES IT IS EXPECTED THAT PERSONNEL USING THIS EQUIPMENT WILL BE COMPETENT WITH THE MANAGEMENT OF APPARATUS WHICH MAY BE POWERED OR UNDER EXTREMES OF TEMPER...

Страница 5: ...vice personnel Contact Isothermal Technology or one of our appointed distributors for details of approved service outlets Operation Before switching on the instrument connect the probe or probes to th...

Страница 6: ...o adjust the requested selection Push ENTER MENU to confirm Push ESC to revert back to the measuring mode MENU Unit Lin2 CAL Chnl Log The calibration is hidden when the instrument has been calibrated...

Страница 7: ...uested calibration option Push ENTER MENU to confirm CAL oFF oP1 oP2 oP3 oP4 1 Standard calibration according IEC 60751 oFF No Corrections are applied Use the up and down keys to select oFF Push ENTER...

Страница 8: ...ing the blinking segment Down key is jumping to the next segment Push ENTER MENU to confirm revert back to the measuring mode Important A break of the physical calibration cannot be done by the ESC bu...

Страница 9: ...p and down keys again to start on or Stop off the logger Push ENTER MENU to confirm Use the up and down keys again to select between automatic storage Auto or manually operated storage SPot Push ENTER...

Страница 10: ...Mode reduced resolution 0 01 from 199 99 to 199 99 Filter Mode high resolution 0 001 from 199 999 to 199 999 Press FAST key to change the measuring rate By using the Filter Mode the instrument is per...

Страница 11: ...llows app 1 hour of further operation The battery symbol indicates according to the battery status between 1 to 3 segments We recommend the use of high capacity alkaline batteries such as the Duracell...

Страница 12: ...o 199 999 C remaining range 0 01 C Accuracy Instrument Only 0 01 C from 80 C to 199 999 0 02 C 0 0015 C reading from 200 C to 660 C Logging Record Average Min and Max over 4000 measurements Measuring...

Отзывы: