background image

                                   

Page 7 of 12 

TTI-10 Iss.02 – 07/11 

The instrument offers different calibration options: 

 

1)

 

 [OFF]: No corrections are applied. (EN 60751) 

 

2)

 

[OP1]: Calibration by code (2 x four digit code) is equivalent to a 2-point calibration. 

 

3)

 

[OP2]: Recommended Mode: Calibration by Comparison to a Standard Thermometer (1-point, 2-point or 3-

 

point calibration)   

 

4)

 

[OP3]: Calibration according to Custom Probe Coefficients (EN60751) 

 

5)

 

[OP4]: Reserved for future use 

 

To Change Calibration Mode

 

 

Push [ENTER/MENU] to calibrate the instrument with sensor. Use the up and down keys [



] to select 

CAL

. Push 

again [ENTER/MENU]. On the left corner of the display appears a small 

1

, which indicates the selected channel. To 

change the channel use the up and down keys [



]. Push [ENTER/MENU] to confirm 

 

 

MENU 

 

 

 

 

 

 

Unit    Lin2   

CAL

*

 

  Chnl    Log   

*

The calibration is hidden when the instrument has been calibrated at Isotech, see page 6 for details.

 

 

Use the up and down keys [



] to select the requested calibration option. Push [ENTER/MENU] to confirm. 

 

 

CAL 

 

 

 

 

 

 

oFF    oP1   

oP2 

  oP3    oP4   

 

1) Standard calibration according IEC 60751 / [oFF 

(No Corrections are applied) 

 

Use the up and down keys [



] to select 

[oFF

. Push [ENTER/MENU] to confirm.  

Push [ESC] to revert back to the measuring mode. 

 

2) Calibration by code / oP1 

 

Use the up and down keys [



] to select 

oP1

. Push [ENTER/MENU] to confirm. 

 

On the bottom of the display appears a small 

1

, after this number a four-digit number (Hex-Code/0..F) is displayed. 

For changing the number use the up key [

]. To step to the next number use the down key [

]. If the requested 

number is complete then push [ENTER/MENU] to confirm. At the bottom of the display a very small 

appears, after 

which a second four-digit number is displayed. For changing the number please follow the manual as before. Push 

[ESC] to revert back to the measuring mode. 

 

Note: 

After confirming 

oP1 

by pushing [ENTER/MENU] the function 

oP1

 (calibration by code) is activated, even 

though you leave the menu by pressing [ESC]. 

 

Display-indication with active calibration code (OP1): 

 

The 

CAL

-segment and the small 

1

 indicates to the user that 

oP1

 is activated. 

 

Содержание TTI-10

Страница 1: ...4 543830 Fax 44 0 1704 544799 Internet www isotech co uk E mail info isotech co uk The company is always willing to give technical advice and assistance where appropriate Equally because of the progra...

Страница 2: ...C F and Ohms 6 Temperature Difference Mode Line2 6 Calibration function CAL 6 Channel Activation Chnl 9 Memory Setup Lo6 9 Recalling the memory data HOLD MAX MIN AVE 10 Measuring rate Normal Mode FAS...

Страница 3: ...PARTS ARE NOT COVERED BY THIS GUARANTEE INTERFERENCE WITH OR FAILURE TO PROPERLY MAINTAIN THIS INSTRUMENT MAY INVALIDATE THIS GUARANTEE RECOMMENDATION The life of your ISOTECH Instrument will be prol...

Страница 4: ...ERSONNEL FAMILIAR WITH GOOD MEASUREMENT PRACTICES IT IS EXPECTED THAT PERSONNEL USING THIS EQUIPMENT WILL BE COMPETENT WITH THE MANAGEMENT OF APPARATUS WHICH MAY BE POWERED OR UNDER EXTREMES OF TEMPER...

Страница 5: ...vice personnel Contact Isothermal Technology or one of our appointed distributors for details of approved service outlets Operation Before switching on the instrument connect the probe or probes to th...

Страница 6: ...o adjust the requested selection Push ENTER MENU to confirm Push ESC to revert back to the measuring mode MENU Unit Lin2 CAL Chnl Log The calibration is hidden when the instrument has been calibrated...

Страница 7: ...uested calibration option Push ENTER MENU to confirm CAL oFF oP1 oP2 oP3 oP4 1 Standard calibration according IEC 60751 oFF No Corrections are applied Use the up and down keys to select oFF Push ENTER...

Страница 8: ...ing the blinking segment Down key is jumping to the next segment Push ENTER MENU to confirm revert back to the measuring mode Important A break of the physical calibration cannot be done by the ESC bu...

Страница 9: ...p and down keys again to start on or Stop off the logger Push ENTER MENU to confirm Use the up and down keys again to select between automatic storage Auto or manually operated storage SPot Push ENTER...

Страница 10: ...Mode reduced resolution 0 01 from 199 99 to 199 99 Filter Mode high resolution 0 001 from 199 999 to 199 999 Press FAST key to change the measuring rate By using the Filter Mode the instrument is per...

Страница 11: ...llows app 1 hour of further operation The battery symbol indicates according to the battery status between 1 to 3 segments We recommend the use of high capacity alkaline batteries such as the Duracell...

Страница 12: ...o 199 999 C remaining range 0 01 C Accuracy Instrument Only 0 01 C from 80 C to 199 999 0 02 C 0 0015 C reading from 200 C to 660 C Logging Record Average Min and Max over 4000 measurements Measuring...

Отзывы: