Page 24 of 43
Low Temp Furnace Model ITL-M-17701 13-07/20
It shows the furnace instrument starting with a setpoint of 250.0°C
•
It is then set to RAMP to 415°C
•
It is then set to DWELL at this temperature for a period of 60 minutes
•
It is then set to RAMP to 421°C over a period of 30 minutes
•
It is then set to DWELL at this temperature for a period of 600 minutes
•
It is then set to STEP to 415°C
•
It is then set to DWELL at this temperature for a period of 600 minutes
•
It is then set to RAMP to 250.0°C
•
It will then END the program and DWELL until switched off
To enable/RUN this PROGRAM:
1. Log in as the SUPERVISOR and exit menu
2. Scroll to the PROGRAM page (Fig5)
3. Press the SCROLL button to highlight the RESET box
4. Press the UP button to highlight the PAGE icon in the top right hand side of the screen
5. Enter this with the SCROLL button
6. This will access the PROGRAM EDIT menu page
7. Scroll to OPERATION and select LOAD, enter with the SCROLL button
8. Scroll to TEST_1 and enter with the SCROLL button
9. The STATUS will display SUCCESS and the CURRENT PROGRAM will display TEST_1
10. TEST_1 is now the default program
11. To RUN the program TEST_1 exit the PROGRAM EDIT page using the PAGE button until you see the
PROGRAM homepage
12. Press SCROLL once to highlight the RESET box
13. Scroll to TEST_1 and press the SCROLL button to initiate the program
The page will now change to show the current program running, as figure 11 above.
To edit an existing PROGRAM:
1. Follow the above procedure to point 10
2. Scroll down to SEGMENT NUMBER and select the SEGMENT you wish to edit using the SCROLL and
UP/DOWN buttons
3. Edit the required parameter (RAMP, DWELL, TIME DURATION etc) using the scroll and UP/DOWN
buttons and enter the value
4. When the editing is complete, it can either be left as the existing PROGRAM or saved under an
alternative file name in the internal memory
5. To leave the alterations under the existing filename simply exit out of the menu and RUN the program
as normal
6. To save as an alternative filename in the memory, SCROLL back up to OPERATION and enter using
the SCROLL button
7. Scroll to STORE and enter
8. The soft keyboard will open allowing you to save the file as a filename of your choice
9. When the name is selected, press the PAGE button to save the changes
10. Both the original file and the modified file will now exist in the memory