5.1.1 Syntax
The commands are transmitted in ASCII. All commands are terminated by the sequence <CR> <LF>
(0x0D 0x0A , 13 10 respectively). Leading zeros can be omitted on input, output is in fixed format.
5.1.2 Command
set
Table 4: Command set
Command Computer
HV-supply
Read module identifier
# *
# * nnnnnn
; n.nn
; U
; I *
(unit number ; software rel.
; V
nom
; I
nom
)
Read break time
W *
W * nnn *
(break time 2 ... 255 ms)
Write break time
W=nnn *
W=nnn * * (break time = 2 - 255 ms)
Read actual voltage channel 1
U1 *
U1 * { polarity / mantisse / exp. with sign } * (in V)
Read actual current channel 1
I1 *
I1 * { mantisse / exp. with sign } *
(in A)
Read voltage limit channel 1
M1 *
M1 * nnn * (in % of V
nom
)
Read current limit channel 1
N1 *
N1 * nnn * (in % of I
nom
)
Read set voltage channel 1
D1 *
D1 * { mantisse / exp. with sign } *
(in V)
Write set voltage channel 1
D1=nnnn.nn *
D1=nnnn.nn * *
(voltage corresponding resolution in V; <M1)
Read ramp speed channel 1
V1 *
V1 * nnn * (2 ... 255 V/s)
Write ramp speed channel 1
V1=nnn *
V1=nnn * * (ramp speed = 2 - 255 V/s)
Start voltage change channel 1
G1 *
G1 * S1=xxx *
(S1 ,
⇒
Status information)
Write current trip
L1=nnnnn *
L1=nnnnn * *
(trip corresponding resolution range mA > 0)
Cannel 1
Range “mA”
LB1=nnnnn *
LB1=nnnnn * *
(trip corresponding resolution range mA > 0)
Range “µA”
LS1=nnnnn *
LS1=nnnnn * *
(trip corresponding resolution range µA > 0)
Read current trip
L1 *
L1=nnnnn * *
(see above, for nnnnn=0
⇒
no current trip)
Cannel 1
Range “mA”
LB1 *
LB1=nnnnn * *
(see above, for nnnnn=0
⇒
no current trip)
Range “µA”
LS1 *
LS1=nnnnn * *
(see above, for nnnnn=0
⇒
no current trip)
Read current trip channel 1
L1 *
L1 * { mantisse / exp. with sign } *
(s.a., current trip in A)
Read status word channel 1
S1 *
S1 * xxx *
(S1 ,
⇒
Status information)
Read module status channel 1
T1 *
T1 * nnn * (code 0...255,
⇒
Module status)
Write auto start channel 1
A1=nn *
A1=nn * *
(conditions
⇒
Auto start)
Read auto start channel 1
A1 *
A1 * nnn * (8
⇒
auto start is active; 0
⇒
inactive)
* = <CR><LF>
The second channel of the supply is addressed by replacing 1 with 2 !
Status information:
ON
Output voltage according to set voltage
OFF
Channel front panel switch off
MAN
Channel is on, set to manual mode
ERR
V
max
or I
max
is / was exceeded
INH
Inhibit signal was / is active
QUA
Quality of output voltage not given at present
L2H
Output voltage increasing
H2L
Output voltage decreasing
LAS
Look at Status (only after G-command)
TRP
Current trip was active