2.2 

Technical Data and auxiliary information 

2.2.1 Device 

class 

 Table 1:   Technical Data, Device class 

TECHNICAL DATA 

HIGH PRECISION 

Single channel HV Power Supply 

SHQ 122 

SHQ 124 

SHQ 126 

Dual channel HV Power Supply 

SHQ 222 

SHQ 224 

SHQ 226 

Output voltage V

nom

 

2 kV 

4 kV 

6 kV 

Output current I

nom 

 

6 mA 

3 mA 

1 mA 

Ripple and noise  

typ.:   < 2 mV

P-P

 

 

max.:   5 mV

P-P

 

Stability: 

 V

O

 /

 V

INPUT

 < 

 10

-5

 (after a warm-up period of 30 min) 

 

V

(no load/ load) 

< 5 

 10

-5

 (after a warm-up period of 30 min) 

Temperature coefficient 

< 3 

 10

-5

/

Voltage 

resolution ADC: 

100 mV / 6-digit LCD display 
(Option

 VHR

10 mV

 for SHQ x22 and x24 

only

measurement accuracy: 

±

 (0,05% V

O

 + 0,02% V

nom

) for one year 

Voltage 

Manual / DAC: 

10-turn potentiometer / digital via serial interface 

Settings resolution 

DAC: 

100 mV / Option 

VHR

SHQ x22M

 with 

30 mV

SHQ x24M 

with 

60 mV

 

Current 

 

2  ranges  / 6-digit LCD display 

measurement 

resolution ADC: 

Range mA: 

 

I

nom

 

 I

O

 

 100 µA, 

Resolution: 100 nA 

Option 

0n1

: I

nom

 

 I

O

 

 10 µA, 

Resolution: 100 nA 

 

 

 

Accuracy: 

±

 (0,1% I

O

 + 0,02% I

nom

 

 

Range µA: 

 

100 µA > I

O

 > 20 nA, 

Resolution: 1 nA  

Accuracy: 

±

 (0,1% I

O

 + 20 nA) 

 

 

 

Option 

0n1

:  10 µA > I

O

 > 2 nA, 

Resolution: 100 pA 

Accuracy:  

±

 (0,1% I

O

 + 2 nA) 

Value scope 

data are guaranteed in the range of (1% 

 V

nom

) < V

O

 < V

nom  

for one year 

Rate of change of 
output voltage 

fixed: 

500 V/s 

(at HV-ON/OFF) 

variable: 

2 ... 255V/s 

(at remote control) 

Protection 

hardware voltage limit   

(V

max

 rotary switch in 10%-steps) 

hardware current limit   

(I

max

 rotary switch in 10%-steps, 

 Option 

IWP

: setting with 10-turn potentiometer 

I

SET

INHIBIT  

 

 

(external signal, TTL, LOW = active) 

programmable current trip  

(serial interface) 

Interface 

RS 232-Interface 

(Option 

CAN

: CAN-Interface 

 SHQ x

4

x) 

Line voltage AC (V

INPUT

 100 V

AC

 .-. 240 V

AC

 / 0,7 – 0,4 A max / 50 – 60 Hz 

Fuse 

 T 1A L 250V / microfuse 5mm x 20mm, 250V / 1AT 

Mains connection 

matching used socket according IEC 60320 C13 (125V/10A or 250V/16A) 

Connectors 

HV output:   V

O nom

 rsp. Typ  

AMP/Tyco SHV-Connector (225059-3) 

INHIBIT:  

V

O max  

 5V  

1-pin Lemo-hub (ERN.00.250.CTL) 

RS 232 :  

V

O max  

+/-20V 

9-pin female Sub-D connector 

Accessories 

see Appendix B 

HV output:   Rosenberger SHV coupler 57K101-006N3 assembled on HV cable  

INHIBIT:  

LEMO FFA.00.250.CTA, standard cable 5V 

RS 232:  

custom comercially available RS232 cable, uncrossed, female to 

 male 

Temperature ranges 

Operating: 

0 ... +40 °C 

Storage:  

-20 ... +60 °C 

Desk case 

 

Size (W/H/D) : (236/100/320) mm 

Содержание SHQ 122

Страница 1: ...k GmbH Email sales iseg hv de Phone 49 0 351 26 996 0 Bautzner Landstr 23 http www iseg hv com Fax 49 0 351 26 996 21 D 01454 Radeberg Rossendorf Germany Operator s Manual High Voltage Power Supply SHQ HIGH PRECISION series ...

Страница 2: ...r The unit may only be operated with protective earth conductor connected We decline all responsibility for damages and injuries caused by an improper use of the module It is strongly recommended to read the operators manual before operation The information in this manual is subject to change without notice We take no responsibility for any error in the document We reserve the right to make change...

Страница 3: ...8 4 Operation 9 4 1 Front panel 9 4 2 Power Up and operational conditions 10 4 2 1 Manual control CONTROL in upper position 10 4 2 2 Remote control CONTROL in lower position 11 4 2 3 Transition manual control Ù remote control 12 4 3 Current measurement ranges 12 4 3 1 Current measurement ranges and current trip 12 4 4 Safety functions 12 4 5 Operation check 13 5 RS232 interface 14 5 1 Specificatio...

Страница 4: ...pplication the unit has to be switched off and discharge of residual voltage has to be finished Depending on application residiual voltages can be present for long time periods These residiual voltages can lead to severe injuries Only accessories which meet the manufacturer s specifications shall be used If the equipment is used in a manner not specified by this manual the protection provided by t...

Страница 5: ...which offer output voltages up to 6 kV for the use in industry and research Main Characteristics High voltage power supplies with front panel operation and remote control via serial interface Output voltages with very low ripple and noise Compact housing with one or two independent high voltage sources Polarity is manually switchable with switches on the rear side of the housing Simultaneous displ...

Страница 6: ... 10 µA Resolution 100 nA Accuracy 0 1 IO 0 02 Inom Range µA 100 µA IO 20 nA Resolution 1 nA Accuracy 0 1 IO 20 nA Option 0n1 10 µA IO 2 nA Resolution 100 pA Accuracy 0 1 IO 2 nA Value scope data are guaranteed in the range of 1 Vnom VO Vnom for one year Rate of change of output voltage fixed 500 V s at HV ON OFF variable 2 255V s at remote control Protection hardware voltage limit Vmax rotary swit...

Страница 7: ...HV filter to provide a feedback voltage for the output voltage control an additional voltage divider supplies the signal for the maximum voltage monitoring A precision control amplifier compares the feedback voltage with the set value given by the DAC remote control or the potentiometer manual control Signals for the control of the resonance converter and the stabilizer circuit are derived from th...

Страница 8: ...ion 3 4 Floating HV outputs Both HV outputs are related to the same ground HV 0V HV GND provided on the outer connector screen of HV cable of SHV connectors The channels can be switched independently in polarity the output voltages related to HV 0V HV GND are also independently controlled The SHV connectors are mounted isolated to chassis PE in order to have a floating HV 0V If the floating voltag...

Страница 9: ...current limit with 10 turn potentiometer The upper figure shows the panel for the SHQ module in the standard version with 10 percent switches for voltage and current limit The front panel for modules with option IWP Hardware current limit setting with 10 turn potentiometer is shown in the lower figure The mains supply including switch and fuse a 9 pin female D Sub connector for the RS 232 interfac...

Страница 10: ...show manufacturer serial number nominal voltage and nominal current After a successful function test the unit is in standby mode If the HV ON switch is in position ON and CONTROL switch is in upper position manual control at the HV output on the rear side high voltage is generated according to the set voltage chosen via the 10 turn potentiometer with a ramp speed of 500 V s hardware ramp After pow...

Страница 11: ...surement of the corresponding SHQ series device The polarity of the output voltage is also indicated 4 2 2 Remote control CONTROL in lower position If the CONTROL switch is in lower position DAC high voltage will be activated only after receiving corresponding RS232 commands There are several commands offering an enhanced control in comparison to front panel control such as voltage ramp The ramp s...

Страница 12: ...r the display In mA position there is an automatic switching between the current measurement ranges in case of crossing of the threshold In position µA current measurement will stay in the µA range in case of measurements higher than the threshold Overflow will be displayed 4 3 1 Current measurement ranges and current trip With the current measurement range switch it is also chosen which current t...

Страница 13: ...ation of the module and the LC display can be verified as follows The CONTROL switches must be in the upper position for manual control For each channel turn the 10 turn potentiometer for VSET by one revolution such that the number in vision panel shows 1 and the circular scale is in position 0 This setting corresponds to 10 of the nominal Voltage ISET must be set to a value 10 No cables are conne...

Страница 14: ...ange of the output voltage 5 1 Specification RS232 interface The data exchange is character based the synchronisation for the transfer direction PC to HV source input is performed using an echo The data transfer to the PC output is asynchronous Between two characters a programmable delay time break is included to allow the computer to receive and evaluate the incoming data The default break time s...

Страница 15: ...nn ramp speed 2 255 V s Start voltage change channel 1 G1 G1 S1 xxx S1 Status information Write current trip L1 nnnnn L1 nnnnn trip corresponding resolution range mA 0 Cannel 1 Range mA LB1 nnnnn LB1 nnnnn trip corresponding resolution range mA 0 Range µA LS1 nnnnn LS1 nnnnn trip corresponding resolution range µA 0 Read current trip L1 L1 nnnnn see above for nnnnn 0 no current trip Cannel 1 Range ...

Страница 16: ...Positive 2 1 4 Negative 0 MAN Control Manual 1 1 2 via RS 232 interface 0 0 0 0 Read out of module status Tx does not reset the flags ERR and INH like the read out of status word Sx of the corresponding channel Auto start Autostart offers the possibility to set a specific voltage which is automatically achieved with a set ramp after Power Up and Initialization of the unit This enables working in r...

Страница 17: ...ay driver Microcontroller Reference voltage ADC MUX DAC 1 DAC 2 Front panel operating elements Resonance Converter HV Transformer Rectifier Polarity switch HV Filter HV OUTPUT Ch1 Channel 1 rectifier AGC amplifier Precision V monitoring max I monitoring max Error logic Inhibit Supply voltage monitoring Hardware voltage ramp Filter HV OUTPUT Ch2 Channel 2 6V 24V Line voltage AC RS 232 Interface DC ...

Страница 18: ...g hv com Fax 49 0 351 26 996 21 18 D 01454 Radeberg Rossendorf Germany Appendix B Accessories Item iseg order number HV cable 5m SHV coupling one sided up to 6kV SHV_CAB5 HV cable 5m SHV coupling both sides up to 6kV SHV2_CAB5 RS 232 cable 3m 200295 LEMO FFA 00 250 CTAC31 200793 Power Supply Cord HG TR SJT3x16AWG C13M 3 05m 595088 ...

Страница 19: ......

Отзывы: