Error codes:
????
Syntax error
?WCN
Wrong channel number
?TOT
Timeout error (with following reinitialization)
?<SP>UMAX=nnnn
Set voltage exceeds voltage limit
Module status:
Status
Description
Bit
Valency
QUA
Quality of output voltage not given at present
7=1
128
ERR
V
max
or I
max
is / was exceeded
6=1
64
INH
INHIBIT signal
was / is active
5=1
32
inactive
0
KILL_ENA
KILL-ENABLE is
on
4=1
16
off
0
OFF
Front panel HV-ON switch in
OFF position
3=1
8
ON position
0
POL
Polarity set to
positive
2=1
4
negative
0
MAN
Control
manual
1=1
2
via RS 232 interface
0
T1:
U/I
Display dialled to
voltage measurement
0=1
1
current measurement
0
Auto start:
Description
Bit
Valency
If module status OFF + ERR + INH + MAN = 0, output voltage of the channel ramping at
set voltage. G-command is not necessary after D-command, POWER-ON and
OFF
⇒
ON.
If output voltage shut off permanently (by ERR or INH at ENABLE KILL or TRP), the
previous voltage setting will be restored with software ramp after
”
Read status word
”
.
3=1
8
Values loading in corresponding
Current trip saving in EEPROM
2=1
4
registers at POWER-ON!
Set voltage saving in EEPROM
1=1
2
Ramp speed saving in EEPROM
0=1
1
(EEPROM guarantee 1 million saving cycles)
Software
Contact us for an overview on our user friendly control and data acquisition software!