SERVICE MANUAL
SYSTEMS
3
1.2) REFRIGERANT SCALE CALIBRATION
Required equipment:
1 Phillips screwdriver, PH 2.
1 Hexagonal spanner no.13.
1 Sample weight, 5 Kg.
2 Sample weights, 2 Kg.
1 Sample weight, 1 Kg.
Or, alternatively:
1 Sample weight, min. 10 Kg.
1 Known weight, 9 Kg.
After removing the front panel, remove the M8 screw that fastens the container (
34
) to the
scale plate (
18
). Remove the container (
34
).
Note: it is not necessary to disconnect the hoses from the internal container (34).
Set menu
0001
(see Chap.1.1), confirm with the
S
Key
.
The display shows
P- - - 0.
With the scale plate (
18
) completely empty, press the
S
Key.
The display shows
P- - - F.
Position the 10-Kg sample weight (or weights) on the scale plate (
18
).
Confirm with the
S
Key
.
The display shows the theoretical weight read by the scale.
Use the
UP-DOWN
Keys to set the sample weight value.
Confirm with the
S
Key.
Remove the 10-Kg sample weight.
The display shows
P- - - t.
Position 9-Kg sample weight (or weights) on the scale plate (
18
).
Confirm with the
S
Key.
The display shows
–End–
.
Press the
S
Key to exit. The display shows the Stand-by mode.
End of procedure.
Содержание Tronic
Страница 11: ...Rev 0128 01 TRONIC SERVICE MANUAL SYSTEMS 11 2 3 REAR VIEW Picture 3 42 2 49...
Страница 15: ...Rev 0128 01 TRONIC SERVICE MANUAL SYSTEMS 15 2 7 POWER BOARD VIEW Picture 7 71 81 80 72 74 75 76 73 79 78 77 70...
Страница 16: ...Rev 0128 01 TRONIC SERVICE MANUAL SYSTEMS 16 2 8 LOGIC BOARD VIEW Picture 8 82 83 85 84...
Страница 17: ...Rev 0128 01 TRONIC SERVICE MANUAL SYSTEMS 17 2 9 FLOW DIAGRAM...
Страница 18: ...Rev 0128 01 TRONIC SERVICE MANUAL SYSTEMS 18 2 10 ELECTRICAL DIAGRAM...
Страница 20: ...Rev 0128 01 TRONIC SERVICE MANUAL SYSTEMS 20 CONTROL PANEL S F UP DOWN...