Doc. SIE10169 Rev. 13 Page 4 of 32
APPLICABLE STANDARDS
The test set conforms to the EEC directives regarding Electromagnetic Compatibility and Low
Voltage instruments.
A) Electromagnetic Compatibility:
Directive no. 2004/108/EC. Applicable Standard : EN61326-1 + A1 + A2.
.
EMISSION
- EN 61000-3-2 + A2: Harmonic content of power supply. Acceptable limits: basic.
- EN 61000-3-3 + A1: Limitation of voltage fluctuations and flicker. Acceptable limits: basic.
- CISPR16 (EN 55011 class A): Limits and measurement methods of radio-electric disturbances for
industrial, medical and scientific instruments at radio-electric frequencies.
Acceptable limits for conducted emission:
. 0.15-0.5 MHz:
79 dB pk; 66 dB avg.
. 0.5-5 MHz:
73 dB pk; 60 dB avg.
. 5-30 MHz:
73 dB pk; 60 dB avg.
Acceptable limits for radiated emission:
. 30-230 MHz:
40 dB (30 m)
. 230-1000 MHz:
47 dB (30 m)
IMMUNITY
- EN 61000-4-2 + A1 + A2: Immunity tests for ESD. Test values: 8 kV in air; 4 kV in contact.
- EN 61000-4-3 + A1; Immunity tests for radio frequency interference. Test values (f= 900
5
MHz): field 10 V/m, modulated AM 80%; 1 kHz
- EN 61000-4-4; Immunity tests for high speed transients (burst). Test values: 2 kV peak; 5/50 ns.
- EN 61000-4-5 + A1; Immunity tests for surge. Test values: 1 kV peak differential mode; 2 kV
peak common mode; 1.2/50 us.
- EN 61000-4-6 + A1: immunity to low-voltage sinusoidal waveform. Test values: 0.15-80 MHz, 10
Vrms, 80% AM 1 kHz.
- EN 61000-4-8 + A1: Immunity tests for low frequency magnetic fields. Test values: 30 Arms/m.
- EN 61000-4-11: Immunity test for power supply drops. Test value: 1 cycle; 100% drop.
B) Low Voltage Directive
- - Directive n. 2006/95/EC. Applicable standard: EN 61010-1. In particular, for a pollution
degree 2:
- Dielectric rigidity 1.4 kV AC, 1 minute.
- Inputs/outputs protection: IP 2X - EN60529.
- Operating temperature: - 10 °C to 55 °C; storage: - 20 °C to + 70 °C.
- Relative humidity : 5 - 95%, without condensing.
- Vibration: IEC 68-2-6 (20 m/s^2 at 10 – 150 Hz);
- Shock: IEC 68-2-27 (15 g; 11 ms; half-sine).
- Altitude: less than 2000 m.
Shop for Power Metering products online at: