background image

4               

 

 

       IRRA

flow

 Laser Leveler User Manual  

7000494 Rev C, USA

 

 

 

 

Scope 

This manual describes the usage of the IRRA

flow

 Laser Leveler model 

number ICLS 010. The IRRAS Laser Leveler is intended to be used with the 

IRRA

flow

 CNS System. The Laser Leveler is used to target the tip of the 

IRRA

flow

 catheter in use.

 

Users  must  read  this  manual  carefully  prior  to  using  the  IRRA

flow

  Laser 

Leveler for the first time so that the features are thoroughly understood. 

Warning - Failure to follow the instructions in this manual may endanger 
the patient and/or the operator! 

 

 

 

 

IRRA

flow

 Laser Leveler Accessory Device 

Description

 

 

The IRRA

flow

 Laser Leveler

 

is a reusable accessory to the IRRA

flow

 CNS 

System and consists of a laser leveler as shown in 

Figure 1 below

.

 

Содержание IRRAflow

Страница 1: ...IRRAflow Laser Leveler 7000494 Rev C USA IRRAflow Laser Leveler REF ICLS 010 User Manual 0344...

Страница 2: ...ls 7 3 1 IRRAflow Laser Leveler Symbols 7 Safety Regulations 8 4 1 General Safety Regulations and Warnings 8 Environmental and Handling Conditions 10 Installation Procedure 10 Instructions for Use 11...

Страница 3: ...activated after pushing on the switch and automatically turns off after 20 seconds IRRAS only accepts responsibility for the equipment s safety usability and performance if the equipment is used in a...

Страница 4: ...the IRRAflow catheter in use Users must read this manual carefully prior to using the IRRAflow Laser Leveler for the first time so that the features are thoroughly understood Warning Failure to follo...

Страница 5: ...Aflow Laser Leveler The laser leveler generates a laser mark to assist in setting the height of the control unit The laser leveler has a power button and auto shuts off after approximately 20 seconds...

Страница 6: ...a product box as show in Figure 2 below This product box includes Laser Leveler Shoulder screw Push on cap Allen wrench Instructions for use The device does not include software It is non sterile can...

Страница 7: ...IRRAflow Control Unit such that the zero point of the Control Unit is at the same elevation as the patient s auditory meatus or ear canal prior to beginning treatment Labels 3 1 IRRAflow Laser Leveler...

Страница 8: ...aindication Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary this equipment and the other equi...

Страница 9: ...n 30 cm 12 inches to any part of the Laser Leveler including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result This is not a toy keep away fr...

Страница 10: ...g IP22 Power Source 3 6V Battery Do not place the device near heating equipment nor expose to direct sunlight Elevated temperatures can result in shortened battery life and degrade performance Install...

Страница 11: ...o the control unit The treatment normally occurs at the bedside but anywhere with direct line of sight to the patient s head may work Rotate the body of the device towards desired target location away...

Страница 12: ...propyl alcohol or 70 ethyl alcohol solution No liquid may be allowed to be directly poured or sprayed onto the Laser Leveler during cleaning or disinfecting as this may damage the equipment No liquid...

Страница 13: ...gnetic environment specified below The customer or the user of the IRRAflow CNS system should assure that it is used in such an environment Standard Description IEC 60601 1 2 4th ed Cispr 11 EN55011 R...

Страница 14: ...4 Rev C USA 10 3 Laser Levelers and Manual Please contact your local distributor if you wish to order Laser Levelers or a replacement manual IRRAflow Laser Leveler Ref no ICLS 010 Laser Leveler User m...

Страница 15: ...ured By Address IRRAS USA Inc 11975 El Camino Real Suite 304 San Diego CA 92130 USA URL http www IRRAs com E mail address US customerservice IRRAs com Phone 1 800 946 0458 EC Representative Address Eu...