Laser Technical Specifications
Wavelength
630-680 nm
Divergence
1.6 mrad
Pulse Pattern
Square
Max Output
< 1 mW
Class
Class 2
10.2.
Electromagnetic Compatibility
Electromagnetic compatibility tests have been performed with a 2.5 m power
cable.
Guidance and manufacturer’s
declaration
–
electromagnetic emissions
The IRRA
flow
Control Unit, and IRRA
flow
Tube Set is intended for use in the
electromagnetic environment specified below. The customer or the user of the
IRRA
flow
CNS system should assure that it is used in such an environment.
Standard
Description
IEC 60601-1-2 4
th
ed
Cispr 11/EN55011
RF Emissions
Group 1, Class B
IEC61000-4-2
ESD
8 kV contact, 15 kV air
IEC61000-4-3
Radiated RF
Immunity
80-2700MHz :3 V/m ordinary, 10V/m home
healthcare, 80% AM, 1 kHz
385MHz :27 V/m, PM,18Hz
450MHz :28 V/m, FM+/-5kHz dev, 1 kHz sine
710, 745, 780MHz :9 V/m, PM, 217 Hz
810, 870, 930MHz :28 V/m, PM, 18Hz
1720, 1845, 1970MHz :28 V/m, PM, 217 Hz
2450MHz :28 V/m, PM, 217 Hz
5240, 5500, 5785MHz :9 V/m, PM, 217 Hz
IEC61000-4-8
Magnetic Immunity
30 A/m, 50 Hz
Table B-1 Electromagnetic compatibility