Cleaning and Disinfection 

The Laser Leveler shall be cleaned and disinfected after each treatment. 

If fluids spill onto the Laser Leveler during treatment, pause or stop treatment 
and wipe off the spillage immediately. 

The recommended method for cleaning is to wipe the parts with a soft cloth 
dampened with soapy water until all foreign material has been removed. 

The recommended method for surface disinfectant is to wipe the parts with 
a  soft  cloth  dampened  with  the  following  intermediate  level  disinfectant 
agents or an intermediate level disinfectant wipe.  

Only the following intermediate level disinfectant liquid may be used with a 
minimum dwell time of one minute exposure: 

70% Isopropyl alcohol or 70% ethyl alcohol solution 

No  liquid  may  be  allowed  to  be  directly  poured  or  sprayed  onto  the  Laser 
Leveler during cleaning or disinfecting as this may damage the equipment. 

No liquid may be allowed to dispel from the cloth during cleaning, as this 
may damage the equipment. 

Never use any type of tools or brush when cleaning, as this may damage the 
equipment. 

Device  is  supplied  non  sterile.  No  components,  parts  or  accessories  to  the 
IRRA

flow

 Laser Leveler may be sterilized.   

IRRAS AB does not make any claims for or representation as to the 
performance characteristics of this product if it is used in conjunction with 
other accessories or other than intended. If there are doubts about how to 
install or clean the Laser Leveler, the effect of cleaning, the functions and/or 
safety of the unit, then the unit should be withdrawn from service and the 
distributor and/or manufacturer consulted (see section 11 for contact 
information). 
 
 
 

 

Содержание IRRAflow

Страница 1: ...IRRAflow Laser Leveler 7000494 Rev C USA IRRAflow Laser Leveler REF ICLS 010 User Manual 0344...

Страница 2: ...ls 7 3 1 IRRAflow Laser Leveler Symbols 7 Safety Regulations 8 4 1 General Safety Regulations and Warnings 8 Environmental and Handling Conditions 10 Installation Procedure 10 Instructions for Use 11...

Страница 3: ...activated after pushing on the switch and automatically turns off after 20 seconds IRRAS only accepts responsibility for the equipment s safety usability and performance if the equipment is used in a...

Страница 4: ...the IRRAflow catheter in use Users must read this manual carefully prior to using the IRRAflow Laser Leveler for the first time so that the features are thoroughly understood Warning Failure to follo...

Страница 5: ...Aflow Laser Leveler The laser leveler generates a laser mark to assist in setting the height of the control unit The laser leveler has a power button and auto shuts off after approximately 20 seconds...

Страница 6: ...a product box as show in Figure 2 below This product box includes Laser Leveler Shoulder screw Push on cap Allen wrench Instructions for use The device does not include software It is non sterile can...

Страница 7: ...IRRAflow Control Unit such that the zero point of the Control Unit is at the same elevation as the patient s auditory meatus or ear canal prior to beginning treatment Labels 3 1 IRRAflow Laser Leveler...

Страница 8: ...aindication Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary this equipment and the other equi...

Страница 9: ...n 30 cm 12 inches to any part of the Laser Leveler including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result This is not a toy keep away fr...

Страница 10: ...g IP22 Power Source 3 6V Battery Do not place the device near heating equipment nor expose to direct sunlight Elevated temperatures can result in shortened battery life and degrade performance Install...

Страница 11: ...o the control unit The treatment normally occurs at the bedside but anywhere with direct line of sight to the patient s head may work Rotate the body of the device towards desired target location away...

Страница 12: ...propyl alcohol or 70 ethyl alcohol solution No liquid may be allowed to be directly poured or sprayed onto the Laser Leveler during cleaning or disinfecting as this may damage the equipment No liquid...

Страница 13: ...gnetic environment specified below The customer or the user of the IRRAflow CNS system should assure that it is used in such an environment Standard Description IEC 60601 1 2 4th ed Cispr 11 EN55011 R...

Страница 14: ...4 Rev C USA 10 3 Laser Levelers and Manual Please contact your local distributor if you wish to order Laser Levelers or a replacement manual IRRAflow Laser Leveler Ref no ICLS 010 Laser Leveler User m...

Страница 15: ...ured By Address IRRAS USA Inc 11975 El Camino Real Suite 304 San Diego CA 92130 USA URL http www IRRAs com E mail address US customerservice IRRAs com Phone 1 800 946 0458 EC Representative Address Eu...

Отзывы: