settings, of course, may be made. The auto mat ic
calibration process is not sacred; it only finds gain
values. The user may change these, if he wishes,
to suit his needs. For example, one may wish to
increase the gain so the nominal displayed value is 
1500 instead of 1200.

Note that the calibration pro ce dure sets the gain of 
the sensor not being used to zero, but one can
adjust the gain of the unused sensor so the S550
responds to both sensors
. The S550 sums the
sensor outputs.

CHECKING OR CHANGING UV GAIN

At the front panel of the P520/P522 press both A
and B push buttons simul ta ne ously. The current
numerical data on the 4 digit display will be
replaced with “CAL?”  meaning, “Do you wish to
Calibrate?” If you press YES you will enter the
Calibration mode described in the pre vious
section. Press NO to enter the Adjust mode. The
display will show “AdJ?” meaning, “Do you wish
to make manual adjust ments?” If you do, press the 
button near YES, which is the B push button.  The
display will show “AuNN” where NN is the
current UV gain with the range 00 - 99. A gain of
00 turns off the sensor. To change the UV gain use
the up or down push buttons. If you then press
YES the new value will be stored; if you press NO
(Channel A push- button) the new value will be
ignored.

CHECKING OR CHANGING IR FILTER

You next have an oppor tunity to change the IR
filter. The display will show “ArFN” where N is
the current filter selected ranging from 1 to 8. Use
the up and down push- buttons to change the filter
if you wish. See 

COMPARISON WITH OTHER

MODEL 500 SERIES VIEWING HEADS

 for a listing

of the high-pass frequencies for these filters. Press 
YES to accept a new filter value or NO to ignore
any changes you may have made.

It is recom mended that the #3 default filter (33Hz
cutoff) be used for all of the initial setup proce -
dures described in this manual. Further infor 

-

mation on how these filters can affect the discrimi -
nation ratio is found in the 

P520 APPLI CATION

MANUAL

 and in the 

P522 APPLI CATION MANUAL

in the section 

MULTI BURNER REQUIRE MENTS.

For further infor mation please call the factory or
contact IRIS Systems Inc. See page 13 for contact
infor mation.

CHECKING OR CHANGING IR GAIN

You now have an oppor tunity to make a final
adjustment of the IR gain. After the filter
selection, the display will show “ANNN” where
NNN, the IR gain, ranges from 000 to 699. The up
and down push- buttons can be used to change the
IR gain. Changed values can be accepted by
pressing YES or ignored by pressing NO. If the
S550 settings are changed, the display will show
“----” indicating that the values are stored in
EEPROM bef ore continuing with the normal
numerical display.

If you do change the IR gain you may be surprised
to see the displayed gain values “jump” as you
scroll through them. As you increase the gain from 
zero you will see this sequence:

000 - 099
150 - 199
250 - 299
350 - 399
450 - 499
550 - 599
650 - 699

The two least significant digits are linear gain
multi pliers; e.g., from 150 to 199 is a 99/50
increase in gain. But the most significant digit is
nonlinear; with each step change in the most
significant digit, the gain doubles. For example, a
setting of 250 has twice the gain as a setting of
150. Now it is possible to see why, as the gain is
increased from 150 to 199 the next step is to 250;
an increase from 150 to 199 is an increase of
99/50, which is a gain increase of almost 2, but the 
next step to 250 is to a gain increase, theore ti cally,
to exactly 2. This relationship is true of all the
gains to 699.

The two least significant digits can be used below
050 down to 000; 000, of course, selects a gain of
zero (which turns OFF the sensor). When the
display is ramped down from 699 the following
sequence will be seen:

PAGE 12

IRIS MODEL S550 APPLICATION MANUAL

Содержание S550

Страница 1: ...PAGE 1 IRIS MODEL S550 APPLICATION MANUAL FLAME MONITORING SYSTEM MODEL S550 S552 S556 VIEWING HEAD APPLICATION MANUAL ...

Страница 2: ... ARE STORED 8 ASSIGNING DIFF S550 SETTINGS TO A B 9 USING TWO VIEWING HEADS PER AMPLIFIER 9 BEFORE MAKING ADJUSTMENTS 10 NEW P520 P522 FRONT PANEL 10 CALIBRATING THE S550 10 EXAMINING RESULTS GAIN SETTINGS CALIB 10 CHECKING OR CHANGING UV GAIN 11 CHECKING OR CHANGING IR FILTER 11 CHECKING OR CHANGING IR GAIN 11 MANUAL SETUP OF THE S550 12 RETURN TO DEFAULT SETTINGS 12 ANGLE OF VIEW 13 GENERAL COMM...

Страница 3: ...Filter Selection 1 8 b IR gain 0 699 When it comes to features like CAL which will be described later the operator is only given those choices which are appropriate For example he does not have to choose between calibrating with the UV sensor or the IR sensor in CAL he is only offered the choice of calibrating with the IR sensor or not This is shown in the FLOWCHART OF S552 PROGRAMMING Figure 4 on...

Страница 4: ...processors Also the earlier IR sensitive family members had only 4 high pass filters to choose from the S550 has 8 S511 S512 S550 LL 36 Hz 1 16 Hz L 71 Hz 2 24 Hz M 105 Hz 3 33 Hz H 186 Hz 4 52 Hz 5 75 Hz 6 100 Hz 7 155 Hz 8 215 Hz TECHNOLOGY USED IN THE S550 The UV and IR sensors in the S550 share a common optical axis efficiently by using a unique dichroic beamsplitting mirror Over 80 of the ult...

Страница 5: ...PAGE 5 IRIS MODEL S550 APPLICATION MANUAL ...

Страница 6: ...e done as follows Remove the plug from its packaging and layout the components 1 Strip 2 inches of the cable outer cover from the cable removing any cellophane wrap or filler material and strip the insulation from the shielded wire if insulated over shield a full 2 inches to expose shield 2 Slide the shield back until a bulge develops close to where the wire exits the cable outer covering 3 Carefu...

Страница 7: ... entirely The reason for this is that there is a complete change in the flicker content for the existing viewing head sighting If this is the case then one should try and optimize the flame signal for the sour gas by zeroing in on this flame not the warm up burner After optimizing the sighting for the sour gas one may find that the signal level is too low on the natural gas One should then try the...

Страница 8: ...ition the zone of higher UV intensity does not overlap the same zones of adjacent or opposing burners so that with proper sighting discrimination is predictable As a general rule the sighting of the S550 when using the UV detector will be more likely a line of sight than the IR flicker detector which will read flame signals bouncing off the side of the sight pipe and other reflectors Sometimes the...

Страница 9: ...n the signal level so if the associated S550 and signal processor has been previously calibrated it will be necessary to calibrate the processor again so that the gain settings will be correct If the discrimi nation is still not satisfactory then try an orificing disc with a smaller diameter hole Generally speaking orificing should always be considered when the burner s are large 50K lbs of steam ...

Страница 10: ...manually changing S550 settings when the desired channel is not selected press the push button for the desired channel and then proceed During the process of examining or changing the S550 settings with the Channel A and B NO and YES push buttons the red LED for the channel which the changes apply will be blinking One should visually check which channel push button is lighted when the S550 setting...

Страница 11: ...ture one does not have the option of working with the channel that is not selected by the external input 3 Bring the firing rate of the burner to a low fire condition 4 If the flame relay has not pulled in adjust the flame on off set points so it does pull in 5 Decide whether to calibrate using the UV or IR sensor See earlier sections on I R DETECTOR and UV DETECTOR At the front panel of the P520 ...

Страница 12: ...r all of the initial setup proce dures described in this manual Further infor mation on how these filters can affect the discrimi nation ratio is found in the P520 APPLICATION MANUAL and in the P522 APPLICATION MANUAL in the section MULTIBURNER REQUIREMENTS For further information please call the factory or contact IRIS Systems Inc See page 13 for contact information CHECKING OR CHANGING IR GAIN Y...

Страница 13: ...tinue reducing the gain in steps of 100 until the displayed reading is reduced Then set the gain so the reading is in the 1000 to 1400 range If the IR sensor was not saturated one can skip paragraph 5 below 5 If the S550 was saturating one should check the aiming of the S550 again using the LED on the back of the housing to adjust the viewing angle of the head for a maximum pulse rate If the LED p...

Страница 14: ...vative product that this is the ultimate state of the art flame monitor The way it works has been described only in a general sense and obviously the algorithms implemented by the embedded controller software are proprietary and will not be made public Any existing P520 or P522 signal processor can be upgraded to an S550 compatible processor by changing the controller chip and the front panel TECH...

Страница 15: ...PAGE 15 IRIS MODEL S550 APPLICATION MANUAL Figure 1 Figure 2 ...

Страница 16: ...IRIS MODEL S550 APPLICATION MANUAL PAGE 16 Figure 3 ...

Страница 17: ...PAGE 17 IRIS MODEL S550 APPLICATION MANUAL Figure 4 ...

Страница 18: ...PAGE 18 IRIS MODEL S550 APPLICATION MANUAL Figure 5 ...

Страница 19: ...PAGE 19 IRIS MODEL S550 APPLICATION MANUAL 5 June 1997 ...

Отзывы: