background image

 

16 

70375-EN Rev B 

577 nm 

Indicated for use in photocoagulation of both anterior and posterior segments including: 

Retinal photocoagulation, panretinal photocoagulation of vascular and structural abnormalities of the 

retina and choroid including:   

Proliferative and nonproliferative diabetic retinopathy 
Choroidal neovascularization 
Branch retinal vein occlusion 
Age-related macular degeneration (AMD) with choroidal neovascularization (CNV) 
Retinal tears and detachments 

Laser trabeculoplasty for the treatment of: 

Primary open angle glaucoma 

Laser iridotomy, iridoplasty for the treatment of: 

Angle closure glaucoma 

Contraindications 

 

Any situation where the target tissue cannot be adequately visualized or stabilized. 

 

Do not treat albino patients that have no pigmentation.  

Potential Side Effects or Complications 

 

Specific to retinal photocoagulation: inadvertent foveal burns; choroidal neovascularization; 

paracentral scotomata; transient increased edema/decreased vision; subretinal fibrosis; 

photocoagulation scar expansion; Bruch’s membrane rupture; choroidal detachment; exudative 

retinal detachment; pupillary abnormalities from damage to the ciliary nerves; and, optic neuritis 

from treatment directly or adjacent to the disc. 

 

Specific to laser iridotomy or iridoplasty: inadvertent corneal or lens burns/opacities; iritis; iris 

atrophy; bleeding; visual symptoms; IOP spike; and, rarely retinal detachment. 

 

Specific to laser trabeculoplasty: IOP spike, and, disruption of the corneal epithelium. 

Specific Warnings and Precautions 

It is essential that the surgeon and attending staff be trained in all aspects of the use of this equipment. 

Surgeons should obtain detailed instructions for proper use of this laser system before using it to perform 

any surgical procedures. 

For additional Warnings and Cautions, refer to Chapter 1, “Introduction.” For more clinical information, 

see “Clinical References” at the end of this chapter. 

Proper eye protection must be utilized for the specific treatment laser wavelength in use (532 nm or 

577 nm).  

Multi-spot mode is intended for retinal photocoagulation only.  

 

For patients with wide variations in retinal pigmentation as evaluated by ophthalmoscopic observation, 

select multi-spot patterns which cover a homogenously pigmented smaller area to avoid unpredictable 

tissue damage. 

Exercise caution while setting multi-spot parameters (pulse duration and the number of spots per pattern) 

when CW laser burns are to be delivered in the macula; with longer grid completion times, the possibility 

of patient movement increases the risk of treatment of unintended targets. 

Содержание TxCell SSLA

Страница 1: ...TxCell Scanning Laser Delivery System Operator Manual...

Страница 2: ...idex logo OcuLight G Probe IQ 532 IQ 577 EndoProbe and MicroPulse are registered trademarks TxCell BriteLight CW Pulse DioPexy EasyFit EasyView FiberCheck IQ 810 LongPulse MilliPulse OtoProbe PowerSte...

Страница 3: ...ENT TREATMENT AND CLINICAL INFORMATION 15 INTENDED USE INDICATIONS FOR USE 15 CONTRAINDICATIONS 16 POTENTIAL SIDE EFFECTS OR COMPLICATIONS 16 SPECIFIC WARNINGS AND PRECAUTIONS 16 PROCEDURAL RECOMMENDA...

Страница 4: ...iv 70375 EN Rev B...

Страница 5: ...nnections The Control Box is paired with an SSLA Cables to connect the SSLA to the Control Box and the Control Box to the laser console Iridex Corporation Contact Information Iridex Corporation 1212 T...

Страница 6: ...eed rate of treatment spot delivery in either single spot repeat or multi spot pattern mode This can result in delivery of laser applications after intended release of the footswitch prior to completi...

Страница 7: ...in the presence of flammables or explosives such as volatile anesthetics alcohol and surgical preparation solutions Turn off the laser before inspecting any delivery device components Always handle th...

Страница 8: ...ket and installation tools depending on the slit lamp model Check the components carefully before use to ensure that no damage occurred during transit Slit Lamp Compatibility Model Spot Size m Slit La...

Страница 9: ...which TxCell SSLA attaches workstation component Fiber optic cable Transmits laser light SSLA Control cable Communicates spot size filter information and scanning information to the control box Contr...

Страница 10: ...er out of the way 3 For Haag Streit equivalent Install mounting bracket or spacer as necessary 4 For Haag Streit equivalent Unlatch ESF from storage position Place SLA on the post of the slit lamp mic...

Страница 11: ...the slit lamp oculars per the images below as applicable 6 Install micromanipulator handle and finger rest as applicable Tighten with thumbscrews ESF Lever Slit Lamp Mounting Platform Setscrew for Se...

Страница 12: ...a minimum loop diameter of 15cm in the fiber optic cable 8 Connect fiber optic cable to the laser console 9 Plug SSLA Control cable into TxCell Control Box The connector will align in a specific orie...

Страница 13: ...ed within the square Both the circle and the square will appear to flash If the observed circle is outside of the square please contact your local Iridex Technical Support representative 5 If aligned...

Страница 14: ...enable laser treatment 2 Move to open position to obtain a clear view unimpeded by a laser Eye Safety Filter NOTE As a safety precaution the laser is unable to enter Treat mode while the Eye Safety F...

Страница 15: ...ntrol knob selects Spacing between spots Grid Triple Arc Circle only The spacing is the distance between the inner edges of a pair of spots Spacing is displayed as increments of spot size diameters an...

Страница 16: ...Grid that is visualized through the slit lamp The projected Target Grid will have a spot centered within one of the cells This spot identifies the size of the associated treatment beam and the cell i...

Страница 17: ...ttern scanning mode B Aiming Beam intensity during single spot mode Automated FiberCheck FiberCheck is an automated test to determine fiber integrity If the fiber needs replacement the following promp...

Страница 18: ...perform single spot test exposure prior to initiating treatment If uncertain of expected clinical response always start with conservative settings and increase laser power and or duration setting in s...

Страница 19: ...s and used to deliver laser energy in CW Pulse MicroPulse or LongPulse mode it is intended to be used by a trained ophthalmologist for the treatment of ocular pathology of both the anterior and poster...

Страница 20: ...irectly or adjacent to the disc Specific to laser iridotomy or iridoplasty inadvertent corneal or lens burns opacities iritis iris atrophy bleeding visual symptoms IOP spike and rarely retinal detachm...

Страница 21: ...he target grid is in sharp focus prior to laser delivery An out of focus target grid may not produce clinically satisfactory results Exposure Duration Heat Flow and Spacing Between Spots When absorbed...

Страница 22: ...e posterior retina Laser treatment parameters may need to be readjusted when treating the peripheral retina Laser Settings It is the physician s responsibility to determine the appropriate treatment p...

Страница 23: ...m Power mW Exposure Duration ms Retinal Photocoagulation 50 1000 50 2000 10 1000 Trabeculoplasty 50 200 500 2000 100 500 Iridotomy 50 200 200 2000 100 300 577 nm MicroPulse Treatment Treatment Spot Si...

Страница 24: ...ch DC Spink CA Subthreshold Diode Micropulse Panretinal Photocoagulation for Proliferative Diabetic Retinopathy Eye Lond 2008 22 5 607 12 10 Muqit MM Marcellino GR Henson DB Young LB Patton N Charles...

Страница 25: ...position If there is still no treatment beam contact your local Iridex Technical Support representative No Pattern Selection button visible on Treatment Screen Turn laser console off Turn Control Box...

Страница 26: ...ue using scanner system E09010 Fan3 speed out of range Warning Chassis fan speed out of range User allowed to continue using scanner system E09011 12V power supply out of range Call Service 12V power...

Страница 27: ...aser did not fire Warning IQ laser did not fire when requested and the current scan pattern has terminated prematurely User is allowed to start a new treatment E09503 Scanner needs calibration Call Se...

Страница 28: ...ce When not in use cover the slit lamp with the provided cover to keep the slit lamp free of dust and store all accessories in suitable storage boxes CLEANING EXTERNAL SURFACES Remove accumulated dust...

Страница 29: ...70375 EN Rev B 25 TO CHECK AND CHANGE SLIT LAMP FUSES...

Страница 30: ...ority to operate the equipment Protection for the Physician Eye safety filters protect the physician from backscattered treatment laser light Integral eye safety filters are permanently installed in t...

Страница 31: ...70375 EN Rev B 27 Labels NOTE The actual label may vary with laser model Serial Number and CE Label Wavelength Label ESF Wavelength Label Haag Streit ESF laser wavelength label...

Страница 32: ...28 70375 EN Rev B Laser Aperture Laser Emission Labels Laser Aperture Laser Emission Labels Control Box Back Label...

Страница 33: ...al with MicroPulse Laser Aperture at End of Fiber Laser Warning Illumination LOT Manufacturer Manufacture Date Off On Part Number Power Pulse Count Pulse Count Reset Non ionizing Electromagnetic Radia...

Страница 34: ...Electrical 100 240 VAC 50 60 Hz Operating temperature range 10 C to 35 C 50 F to 95 F Storage temperature range 20 C to 60 C 4 F to 140 F Maximum recommended ambient air temperature for treatment 30 C...

Страница 35: ...cy range The wireless footswitch has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable p...

Страница 36: ...environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors shoul...

Страница 37: ...in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m a Fields strengths from fixed RF transmitters as determined by an electromagnetic site sur...

Страница 38: ...er of Transmitter W Separation Distance According to Frequency of Transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1...