EMC Safety Information
The laser system (console and accessories) needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in this section. Portable and
mobile RF communications equipment can affect this system.
This laser system has been tested and found to comply with the limits for medical devices in IEC
60601-1-2 according to the tables in this section. These limits are designed to provide reasonable protection
against harmful interference in a typical medical installation.
CAUTION:
Changes or modifications to this laser system not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment and may result in increased
emissions or decreased immunity of the laser system.
The wireless footswitch transmits and receives in the frequency range of 2.41GHz to 2.46GHz with a
limited effective radiated power as described below. The transmissions are continuous transmissions at
discrete frequencies within the transmission frequency range.
The wireless footswitch has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If the wireless footswitch does cause harmful interference to radio or television
reception, which can be determined by turning the laser system off and on, the user is encouraged to try to
correct the interference by one or more of the following measures:
• Reorient or relocate the receiving device.
• Increase the separation between the equipment.
• Connect the laser console into an outlet on a circuit different from that to which the receiver is
connected.
• Consult Iridex Customer Service for help.
This Class B digital apparatus meets all requirements of the Canadian Interference-Causing
Equipment Regulations.
Cet appareil numérique de la classe B respecte toutes les exigences du Réglement sur le matériel
brouilleur du Canada.