Connecting the Power Supply and Air Pump
The 1625 laser comes with a separate power supply and air pump unit.
The unit also contains a long life air pump and filters designed to provide exactly the
right flow and pressure of clean air to the laser window. The laser will not operate if the
air flow is not set
correctly. It is NOT recommended that “house air” be used because of
the exacting air pressure limits and flow limits that are monitored internally.
Pressure
above 1.5 psi (0.1 Bar) can cause damage. Connect the supplied blue air hose to the
laser and power supply air connection points. The blue hose is made of a material that
does not emit plasticizers in vapor form that can damage the window over time. Do not
use other hoses without checking with Iradion.
A twist lock plug is provided that will wire to the AC power source. This plug is inserted
in the twist lock power interface. The Breaker must be OFF when power is applied to
avoid powering up the laser accidentally.
Position the power supply unit in the host system such that access to the air filters is
convenient. Frequent servicing of the filters may be required depending on how clean
and dry the air is around the unit.