E04
BNC electrode slope error.
E04 is displayed if the electrode
slope is less than 85% of the
theoretical maximum of 59.16mV
per unit of pH.
A slope error indicates poor pH
electrode condition. Follow
cleaning instructions.
IF E04 persists, replace electrode.
Override E04 by pressing
pH/mV
key for temporary use of the
electrode.
E05
Wrong Buffer.
Place probe in correct buffer.
Follow calibration instructions in
this manual.
Dirty sensor may slow response
time so that the buffer is
incorrectly identified. Clean
sensor and recalibrate.
If you are confident that the
buffer is correct, override E05 by
pressing the
CAL
key.
E06
ISFET slope out of range.
Follow the cleaning instructions
Calibration error.
Carefully follow the two point
calibration instructions.
Damaged probe
If problem persists, replace probe.
E07
ISFET offset out of range.
Stir the probe in solution to
remove any air bubble trapped on
the sensor surface.
If problem persists, replace probe.
E09
Damaged ISFET sensor.
Replace probe.
14.00 or
0.00
Damaged sensor or probe
connection problem.
See Troubleshooting guide.
Flashing
PROBE
Damaged ISFET probe
advance warning.
Slightly damaged sensor surface.
Probe is still usable. Replacement
probe should be kept on hand.