MODEL IQ150 pH Meter
Getting Started
TWO-POINT CALIBRATION--QUICK REFERENCE GUIDE
1.
2.
3.
4.
Connect probe to
meter.
Put probe in first
calibrating buffer.
Press
ON/OFF
to
turn on meter.
If necessary, press
pH
until pH icon is
seen on the display.
5.
6.
7.
8.
Press
CAL
.
If necessary, press
SELECT BUFFER
until display matches
first buffer value.
Press
ENTER
and
wait until the large
number display
stops flashing.
When complete,
meter will display
the next stored
buffer value.
2PT
will be displayed.
9.
10.
11.
12.
Rinse probe in
deionized water
and place in
second calibrating
buffer.
If necessary, press
SELECT BUFFER
until display matches
second buffer value.
Press
ENTER
and
wait until the large
number display
stops flashing.
Rinse probe in
deionized water and
place in sample.