![IQ IQ150 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/iq/iq150/iq150_instructions-manual_2089449019.webp)
18
Cleaning with Pepsin (Protein Deposit Removal)
Digest proteins by soaking for 15 minutes to one hour in Probe Cleaning
Solution, available from IQ Scientific Instruments as item #PCS50. Rinse
well with deionized water. Soak the probe for at least two hours in pH 4.00
buffer.
Etching with Ammonium Biflouride (DO NOT ETCH ISFET PROBES)
This is generally considered the method of last resort.
If the BNC electrode
response is still sluggish after trying other cleaning methods, immerse
electrode tip in 0.1M NH
4
HF
2
(ammonium biflouride) for one minute. This
is a solution of about 5.7 grams ammonium biflouride to 1 liter of distilled
water. This will etch the surface of the glass and may restore performance.
Rinse well with deionized water. Soak the probe overnight in pH 4.00 buffer.
CAUTION:
This solution is extremely hazardous.
Protective clothing and eyewear must be worn.
Dispose of used solution properly.
Titrations
Should you wish to use the IQ150 pH “Dual Technology” meter in a titration
application, please contact IQ Scientific Instruments for instructions.