TIGER XTS Instrument User Manual V1.0
P
ion
eering Gas Sensing Technology.
Page 39 of 74
ionscience.com
If required, tick
Zero on Carbon Canister
to ensure a near Zero reading after calibration. The
should also be selected.
Tiger PC offers two-point (Zero + SPAN 1) and three-point (Zero + SPAN 1 + SPAN 2) calibration
options:
•
For two-point calibration, enter the SPAN 1 concentration and ensure that the SPAN 2
checkbox is unticked.
•
For three-point calibration, tick the checkbox and enter the SPAN 1 and SPAN 2
concentrations.
Select
Send to instrument
to save these settings to your TIGER XTS.
The calibration procedure is detailed in
Calibration Reminder
The calibration reminder allows you to set a reminder for when calibration is needed. To turn on the
calibration reminder, ensure the box is ticked next to ‘calibration reminder’. You then have two
options below that. You can either set how many days, weeks or months you want the reminder to
appear after the day you are on, or you can set a more precise date by selecting the bottom option.
The calibration reminder looks like the below.
OK
Calibration overdue
7/4/2022