The cleaning procedure is as follows:
1.
Open the vial of Aluminium Oxide polishing compound. With a clean cotton bud collect a
small amount of compound.
2.
Use this cotton bud to polish the Lamp detection window. Use a circular action applying light
pressure to clean the detection window. Never touch the detection window with fingers.
3.
Continue polishing until an audible “squeaking” is made by the cotton bud with compound
moving over the window surface (usually within fifteen seconds).
4.
Remove the residual powder with a short blast of air from the can of clean, dry, oil free air.
5.
Refit the Lamp into the Electrode Stack as described previously.