P 1, 2, 3 & 5000 Ex Ion Science Ltd
Page 9 of 52
Getting started
1. Turn the instrument on by pressing the
key.
2. Press and hold the
key and check the correct gas, alarm levels and zero mode are selected.
3. The screen will flash for 1 minute while the detector warms up – check for a reading on the display
Between 0.2ppm and 2ppm.
4. Replace the probe and filter with the blue carbon filter – check for a reading from 0.0 to 0.2ppm.
5. Remove the carbon filter and replace the probe and filter.
6. If the results are not as expected, perform a custom calibration first ensuring the correct gas is
selected.
Setting alarm levels
We recommend that alarm levels are set to user specifications as soon as possible before being used for the
first time. Please see page 19 for details of how to set alarm levels.
Setting date and time
We recommend that the correct local date and time be set to the instrument as soon as possible before
being used for the first time. Please see page 23 for details on how to set the date and time.
Selecting a gas from the inte rnal gas table
P instruments leave Ion Science pre-set for gas type TVOC (total volatile organic compounds).
Instruments are factory calibrated against isobutylene and all response factors are equivalent to this. By
changing the gas from the gas table all readings will be given using that gases response factor. To change
the selected gas please see pages 23 & 24.
Rechargeable Batteries
Although P instruments leave the factory fully charged, prolonged periods of storage may result in
the battery pack discharging. We recommend charging the instrument for 12 hours before use. See the
battery section of this manual page 38.
Alkaline Batteries
Fit the battery pack supplied with the P. For more information please see the battery section of
this manual page 38.