About Hydrosteel 6000   

 

 

 

 

 

 

 

 

 

 

Principle of Operation 

 
Collecting a hydrogen sample 
Hydrosteel  6000  enables  measurements  to  be  made  of  gaseous  hydrogen  flux  emanating  from  a  steel 
surface. The steel surface can be of variable curvature and surface condition.  
 
To make the measurement, a stream of ambient air is drawn across the steel surface by a high performance 
low-power-demand  pump.  This  test-

gas  stream  is  confined  within  a  specially  designed  ‘collector  plate’ 

contained  in  the  probe  assembly  which  is  attached  to  the  steel  surface. The  probe  assembly  and  collector 
plates vary according to application. The probe types are listed with their features in 

Section 

– Application 

specific measurements.

  

 
The collected hydrogen is conveyed through narrow bore tubing, into the measurement analyser, and across 
the face of a very sensitive amperometric detector. 
 
Determining the level of the collected hydrogen 
The  zero  hydrogen  datum  point  for  Hydrosteel  6000  measurement  is  the  background  hydrogen 
concentration in air at sea level, namely 0.5 ppm

1

.  Rapid diffusion, and hence rapid dispersion of hydrogen 

in air, results in a very low and stable background prevailing in the ambient air of most environments, even 
indoors. This condition enables the Hydrosteel 6000 to be used in measuring a very small flux of hydrogen 
emanating  from  steel  as  a  result  of  its  interaction  with  hydrogen  occluders  and  hydrogen  occluding 
processes. The integrity of the measured data is circumscribed by drift in the ambient air hydrogen content, 
and this drift is usually very low.  
 
The flow of the test-gas stream 

across the steel surface is carefully regulated by means of restrictions, and 

a  flow  bypass,  between  the  pump  and  the  detector  to  ensure  that  it  remains  smooth  and  constant.    An 
additional  feature  of  the  flow  bypass  is  to  ensure  that  in  the  event  of  a  probe  blockage,  the  pump  is  not 
working  against  a  negative  pressure  differential.    Flow  regulation  is  particularly  important  as  the  flux 

J

  of 

hydrogen emanating from the steel is given by: 

J

 = c x 

/ A 

 
where 

A

 is the effective area over which hydrogen is captured, and 

c

 the enhanced hydrogen concentration 

resulting from hydrogen entrainment.   

 

Output  from  the  amperometric  sensor  is  an  electric  current,  which  is  proportional  to  the  concentration  of 
hydrogen presented to it. This current is amplified and scaled for display, and logging, as a hydrogen flux. 
 
To  summarise, the  prowess  of  Hydrosteel  6000  technology  stems from  capturing  the  hydrogen  flux  from  a 
well-defined  area  of  steel  surface  in  a  known  flow  of  air,  and  measuring  the  very  small  enhancement  in 
hydrogen concentration entrained in the air stream.  By this means, the very small hydrogen flux commonly 
emanating from steel can be quantitatively measured, easily, rapidly and reliably. 
 

Application specific measurements 

Hydrosteel  can  be  used  to  locate,  map,  and  monitor  hydrogen  flux  as  detailed  in 

Section 

–  Deployment 

Methods 

on  the  next  page.  Hydrosteel  engages  interchangeable  probes  to  enable  attachment  to  steel  of 

diameter  greater  than  2  in.,  5  cm,  and  surface  temperatures  from  -40  to  +500  ºC,  -40  to  930  ºF.        The 
interchangeable probes are classified as follows: 

 
LT-R: Low Temperature 

– Roaming Probe 

This  probe  utilises  a  collector  plate  6  in.,  15  cm  across  to  deliver  the  maximum  sensitivity,  and  is 
magnetically attached to steel of 3.5 in., 9 cm diameter or greater, of surface temperatures up to 130 

o

C, 265 

o

F  for  short  periods  (quick  spot  measurements  only),  and  110 

o

C,  230 

o

F  for  general  use  (for  extended 

periods please consider use of AT-S probe). To ensure efficient hydrogen capture in wet or windy conditions, 
the LT-R probe incorporates a flexible seal around its perimeter. 

 

 
 
 
 

Содержание Hydrosteel 6000

Страница 1: ...Part number 40010 Unrivalled Detection www ionscience com Hydrosteel 6000 Instrument User Manual V2 1...

Страница 2: ...t The standard warranty of your instrument can be extended to up to five years on PhoCheck Tiger and two years on other Ion Science instruments To receive your extended warranty you must register your...

Страница 3: ...N 13908 2002 ATEX Application of quality systems BS EN 50014 1998 ATEX general requirements BS EN 50020 2002 ATEX Intrinsic safety i BS EN 61010 1 2010 Safety requirements for measurement control lab...

Страница 4: ...ity 8 Initial Appraisal 9 System Assembly and Start up 9 Using Hydrosteel 6000 10 Analyser Operation Menu 13 Analyser display 14 Guide to Field Measurements 16 Data Logging and Computer Interface 19 U...

Страница 5: ...equipment should be used in accordance with this manual and in compliance with local safety standards Legal Notice Whilst every attempt is made to ensure the accuracy of the information contained in...

Страница 6: ...A IIB and IIC are intermittently present zone 1 with an ambient temp range of 20 C to 50 C Intrinsically safe installations will require the use of safety barriers and appropriate wiring b Non sparkin...

Страница 7: ...e pump and the detector to ensure that it remains smooth and constant An additional feature of the flow bypass is to ensure that in the event of a probe blockage the pump is not working against a nega...

Страница 8: ...analyser has a real time display of flux which can be used to search for sites of hydrogen permeation Since the analyser commences measurement at the time of attachment to a surface and a 50 response...

Страница 9: ...ter case 3 Using the special key provided open the battery compartment at the rear of the analyser and insert the four batteries complete with their holder supplied see the diagram below Close and loc...

Страница 10: ...he connector at the top of the analyser as shown below Tighten finger tight Note we recommend that you use the shorter of the two sample tubes However if you require the extension arms you may find it...

Страница 11: ...r assembly from the box and remove the magnet sleepers 2 Manually unscrew the collector handle from the assembly The handle unscrews below the white heat shield 3 Insert the HT R 2ft flexible conduit...

Страница 12: ...remove the magnet sleepers 2 Manually unscrew the collector handle from the assembly The handle unscrews below the white heat shield 3 Insert the HT R very HT extension conduit assembly A 41017 throug...

Страница 13: ...cool dry place Analyser Operation Menu Three front panel keys On the analyser front panel the three keys S Spanner D Data and O On Off have dual functions according to whether the analyser is in star...

Страница 14: ...1 nl cm 2 s When the collector is not attached to a steel surface a flux continues to be displayed in order to give a clear indication of any background drift which may affect the integrity of subsequ...

Страница 15: ...ng HT HT R high temperature roaming blank ambient hydrogen measurement probe independent Using an External Power Supply To extend the continuous operation time of the analyser particularly for long te...

Страница 16: ...meter high temperature sample tube in addition to the thermal jacket this will allow the analyser to be shielded and kept at a greater distance from the hot test area Connect the appropriate probe Aft...

Страница 17: ...3 inch diameter high temperature probe Introduce the probe directly to the steel surface Any orientation of the probe is acceptable but take particular care with piping of less than 12 inches diameter...

Страница 18: ...ign with the IRDA link box and execute the file ISPLC EXE Make sure test equipment is clean and dry return to the field case and store field case in a cool dry location WARNING The HT R probe and asso...

Страница 19: ...access short data logging by pressing the D data key twice While attach probe is displayed attach the probe to the target steel surface The memory bar flashes at one second intervals 2 Twelve 5 secon...

Страница 20: ...aders to instrument Checking instrument details eg serial number logged memory used Displaying and recording flux measurements on PC in real time Re programming analyser with a firmware upgrade Hardwa...

Страница 21: ...s serial port that your IR Link box is connected to To select a port enter the ISLPC software program and click on the Settings tab Downloading Data from your Hydrosteel 1 Remove the Hydrosteel analys...

Страница 22: ...If the communication with the analyser fails the message window shown in step 6 will be displayed Data download display Data files downloaded from an instrument are presented in three windows 1 Data...

Страница 23: ...o reference and Log Linear scale required for the recorded response Using the cursor Right Click on the graphical analysis window to reveal the dropdown menu with the following options Elapsed Real ti...

Страница 24: ...r the file validity before opening data and VALID will then be inserted into the header data of an open file Close Closed highlighted Numerical or Graphical window Caution Before closing the programme...

Страница 25: ...create a set of zone headings Enter up to 60 characters into each of the 199 zone addresses These can then be downloaded to Hydrosteel Having created the headings you can also save them as type zon f...

Страница 26: ...and LT R probe plates are only liable to damage and degradation after extensive use when attached to a steel surface having ridges or high spots Damage will be evident from worn or crushed grooves on...

Страница 27: ...ation Flow test The Hydrosteel flow is important to the accuracy of flux measurements The function test uses a variable area flow meter to check the correct flow is being drawn into the instrument The...

Страница 28: ...need to be returned to ISL for service and calibration Sticking floats There is limited clearance between the float and the inside of the tube In some cases flow tubes are found to stick after transpo...

Страница 29: ...1 Switch unit on and let it complete the stabilisation count down and zero Note the probe type indicated by the analyser display LT or HT 2 Ensure that the temperature reading in the bottom right hand...

Страница 30: ...tube to the to the regulator 6 Press the barb of the flood leak adapter into the other end of the tube 7 Check the pressure on the regulator is not zero If the gas bottle becomes empty then please co...

Страница 31: ...cated 2 Connect the Hydrosteel 6000 analyser to the 2 ft LT R sample tube then connect the end of the sample tube to the flood leak as shown 3 Leave the instrument connected for 1 to 2 minutes for the...

Страница 32: ...of in the same way that it does when the batteries die 1 Using the battery screwdriver provided remove the battery cover at the underside of the analyser 2 Remove the exhausted batteries from the bat...

Страница 33: ...ector plates The solvent vapour in strong concentrations may affect the sensor readings Attention must be made to keep the following areas clean 1 All pneumatic joints 2 Analyser inlet 3 Surface of co...

Страница 34: ...ence com instrument registration Service At Ion Science we recommend that all of our gas detection instruments be returned for service and factory calibration once every 12 months Contact Ion Science...

Страница 35: ...nts Batteries 4 x 1 5 V AA alkaline Time operable with batteries 30 hr occasional 40 hr continuous Monitoring Environmental operating conditions Ambient temperature range 15 F to 120 F 10 C to 50 C Te...

Страница 36: ...cm 2 s Resolution 2 pl cm 2 s hydrogen emanating from steel Linearity Linear over full range Reproducibility 5 or 1 pl cm 2 s whichever is greater Rate of response FSD 1 50 90 27 inch 70 cm conduit 8...

Страница 37: ...the prompt start up mode is displayed press D Display data log summary 10 s D DD Data download instructions 10 s D DDD Data download 30 s D DDDD Cancel download and return to Start up mode none Also d...

Страница 38: ...a log to 10 min data intervals Select D to end logging inde finite D low battery log full D DS Access log statistics file length maximum flux slope of present flux end log mode and return to operating...

Страница 39: ...T S Probes 13 01 09 V2 38 V3 25 Hydrosteel 6000 to V1 9 Warranty added on cover Declaration of Conformity updated Contents updated Warranty and Service info added pg 31 23 07 10 V2 38 V3 25 Hydrosteel...

Отзывы: