Guide to field measurements 

 

 

 

 

 

 

 

 

 

 
Step 1:  Preparing the equipment for flux measurements. 

Open the field case.  Connect the hand held analyser to either the 6” diameter LT probe in the field case or 
the HT probe in the HT-probe kit box, using the appropriate gas sampling tube. This comprises essential test 
equipment,  it  weighs  1  kg,  and  can  be  conveniently  carried  in  a  shoulder  bag  or  small  rucksack.  Before 
commencing a prolonged series of spot measurements, or extended monitoring at one particular site, ensure 
sufficient battery power is indicated  by a full battery indicator on the analyser display.  If necessary replace 
batteries using the battery tray key. 
  

Step  2:  Confirming  suitability  of  ambient  conditions. 

Turn  on  the  analyser  (I/O  key)  in  the  open  air  at 

least 10 m from any source of hydrogen. If  you are using a different probe than previously, reconfigure the 
analyser during the normalisation countdown (see 

Appendix B

 for menu) to indicate the correct probe (LT - 

low  temperature  or  HT 

–  high  temperature).  The  analyser’s  internal  temperature  will  be  indicated  on  the 

display  panel.  A  flashing  reading  indicates  that  the  analyser  is  experiencing  an  abrupt  change  in 
temperature, in  which  case  allow  several  minutes  for  the  flashing  to  subside.    It  is  highly  recommended  to 
use  the  thermally  insulating  leather  jacket  to  reduce  the  effect  of  abrupt  temperature  changes.  Abrupt 
temperature  changes  may  occur  for  many  reasons.    Most  common  are  moving  from  a  heated  or  air 
conditioned  environment  to  the  cold  or  hot  local  conditions.  Alternatively  the  radiant  heat from  hot  pipes  or 
equipment may create a significant temperature change.  When approaching hot pipes use the 2 meter high 
temperature  sample  tube  (in  addition  to  the  thermal  jacket)  this  will  allow  the  analyser  to  be  shielded  and 
kept at a greater distance from the hot test area. Connect the appropriate probe.   
 
After  the  countdown  the  analyser  will  display  a  reading  of  less  than  5  pl/cm

2

/s with  the  LT  probe  and  0.03 

nl/cm

2

/s with the HT probe.   

 
Approach  the  test  site  location  of  interest.  The  only  gas  liable  to compromise  dependable  measurement  of 
hydrogen flux is hydrogen itself, for example

 

within a few metres of hydrogen leaking from a flange.  This will 

be  registered  on  the  analyser  display  and  in  the  analyser’s  memory,  enabling  confidence  in  the  logged 
measurements to be calculated when data is downloaded to a PC.  
 

Step 3: Steel test surface site selection and preparation. 

  

With the LT-R (6 inch diameter low temperature) probe: 

7 inch diameter test surface sites should be identified on mild steel vessels or piping of 3.5 inches diameter 
or  greater,  and  of  surface  temperature  of  less  than150 

o

C,  300 

o

F.    Locate  sites  which  are  smooth, 

particularly avoiding ridges or grooves.   
 
Remove  dust,  loose  rust  and  liquids  and  disbonded  paint  work  from  the  site.    Disbonded  paint  is  usually 
evident from flaky or cracked paint in the locale of the target site, and should be completely removed. 

 
With the HT-R (2.3 inch diameter high temperature) probe: 

3  inch  diameter  test  sites  should  be  identified  on  carbon  or  low  alloy  steel  vessels  or  piping  of  8  inches 
diameter  or  greater,  and  of  surface  temperature  of  less  than  500 

o

C,  930 

o

F.    For  tests  through  inspection 

ports, ensure that there is a clear pathway 
through  the  port  to  the  steel  surface,  so  that  when  the  probe  is  inserted,  it  engages  fully  with  a  3  inch 
diameter  surface  of  steel,  without  the  intervention  of  insulation  wool.  Ensure  that  the  test  surface  is 
completely free of any layers of rust.  
 

Step 4: Data handling.

  If required, access the test site zone menu and data logging menu by pressing the 

keypad 

Data

 key. 

 

 

 
 

 
 
 
 
 
 

Содержание Hydrosteel 6000

Страница 1: ...Part number 40010 Unrivalled Detection www ionscience com Hydrosteel 6000 Instrument User Manual V2 1...

Страница 2: ...t The standard warranty of your instrument can be extended to up to five years on PhoCheck Tiger and two years on other Ion Science instruments To receive your extended warranty you must register your...

Страница 3: ...N 13908 2002 ATEX Application of quality systems BS EN 50014 1998 ATEX general requirements BS EN 50020 2002 ATEX Intrinsic safety i BS EN 61010 1 2010 Safety requirements for measurement control lab...

Страница 4: ...ity 8 Initial Appraisal 9 System Assembly and Start up 9 Using Hydrosteel 6000 10 Analyser Operation Menu 13 Analyser display 14 Guide to Field Measurements 16 Data Logging and Computer Interface 19 U...

Страница 5: ...equipment should be used in accordance with this manual and in compliance with local safety standards Legal Notice Whilst every attempt is made to ensure the accuracy of the information contained in...

Страница 6: ...A IIB and IIC are intermittently present zone 1 with an ambient temp range of 20 C to 50 C Intrinsically safe installations will require the use of safety barriers and appropriate wiring b Non sparkin...

Страница 7: ...e pump and the detector to ensure that it remains smooth and constant An additional feature of the flow bypass is to ensure that in the event of a probe blockage the pump is not working against a nega...

Страница 8: ...analyser has a real time display of flux which can be used to search for sites of hydrogen permeation Since the analyser commences measurement at the time of attachment to a surface and a 50 response...

Страница 9: ...ter case 3 Using the special key provided open the battery compartment at the rear of the analyser and insert the four batteries complete with their holder supplied see the diagram below Close and loc...

Страница 10: ...he connector at the top of the analyser as shown below Tighten finger tight Note we recommend that you use the shorter of the two sample tubes However if you require the extension arms you may find it...

Страница 11: ...r assembly from the box and remove the magnet sleepers 2 Manually unscrew the collector handle from the assembly The handle unscrews below the white heat shield 3 Insert the HT R 2ft flexible conduit...

Страница 12: ...remove the magnet sleepers 2 Manually unscrew the collector handle from the assembly The handle unscrews below the white heat shield 3 Insert the HT R very HT extension conduit assembly A 41017 throug...

Страница 13: ...cool dry place Analyser Operation Menu Three front panel keys On the analyser front panel the three keys S Spanner D Data and O On Off have dual functions according to whether the analyser is in star...

Страница 14: ...1 nl cm 2 s When the collector is not attached to a steel surface a flux continues to be displayed in order to give a clear indication of any background drift which may affect the integrity of subsequ...

Страница 15: ...ng HT HT R high temperature roaming blank ambient hydrogen measurement probe independent Using an External Power Supply To extend the continuous operation time of the analyser particularly for long te...

Страница 16: ...meter high temperature sample tube in addition to the thermal jacket this will allow the analyser to be shielded and kept at a greater distance from the hot test area Connect the appropriate probe Aft...

Страница 17: ...3 inch diameter high temperature probe Introduce the probe directly to the steel surface Any orientation of the probe is acceptable but take particular care with piping of less than 12 inches diameter...

Страница 18: ...ign with the IRDA link box and execute the file ISPLC EXE Make sure test equipment is clean and dry return to the field case and store field case in a cool dry location WARNING The HT R probe and asso...

Страница 19: ...access short data logging by pressing the D data key twice While attach probe is displayed attach the probe to the target steel surface The memory bar flashes at one second intervals 2 Twelve 5 secon...

Страница 20: ...aders to instrument Checking instrument details eg serial number logged memory used Displaying and recording flux measurements on PC in real time Re programming analyser with a firmware upgrade Hardwa...

Страница 21: ...s serial port that your IR Link box is connected to To select a port enter the ISLPC software program and click on the Settings tab Downloading Data from your Hydrosteel 1 Remove the Hydrosteel analys...

Страница 22: ...If the communication with the analyser fails the message window shown in step 6 will be displayed Data download display Data files downloaded from an instrument are presented in three windows 1 Data...

Страница 23: ...o reference and Log Linear scale required for the recorded response Using the cursor Right Click on the graphical analysis window to reveal the dropdown menu with the following options Elapsed Real ti...

Страница 24: ...r the file validity before opening data and VALID will then be inserted into the header data of an open file Close Closed highlighted Numerical or Graphical window Caution Before closing the programme...

Страница 25: ...create a set of zone headings Enter up to 60 characters into each of the 199 zone addresses These can then be downloaded to Hydrosteel Having created the headings you can also save them as type zon f...

Страница 26: ...and LT R probe plates are only liable to damage and degradation after extensive use when attached to a steel surface having ridges or high spots Damage will be evident from worn or crushed grooves on...

Страница 27: ...ation Flow test The Hydrosteel flow is important to the accuracy of flux measurements The function test uses a variable area flow meter to check the correct flow is being drawn into the instrument The...

Страница 28: ...need to be returned to ISL for service and calibration Sticking floats There is limited clearance between the float and the inside of the tube In some cases flow tubes are found to stick after transpo...

Страница 29: ...1 Switch unit on and let it complete the stabilisation count down and zero Note the probe type indicated by the analyser display LT or HT 2 Ensure that the temperature reading in the bottom right hand...

Страница 30: ...tube to the to the regulator 6 Press the barb of the flood leak adapter into the other end of the tube 7 Check the pressure on the regulator is not zero If the gas bottle becomes empty then please co...

Страница 31: ...cated 2 Connect the Hydrosteel 6000 analyser to the 2 ft LT R sample tube then connect the end of the sample tube to the flood leak as shown 3 Leave the instrument connected for 1 to 2 minutes for the...

Страница 32: ...of in the same way that it does when the batteries die 1 Using the battery screwdriver provided remove the battery cover at the underside of the analyser 2 Remove the exhausted batteries from the bat...

Страница 33: ...ector plates The solvent vapour in strong concentrations may affect the sensor readings Attention must be made to keep the following areas clean 1 All pneumatic joints 2 Analyser inlet 3 Surface of co...

Страница 34: ...ence com instrument registration Service At Ion Science we recommend that all of our gas detection instruments be returned for service and factory calibration once every 12 months Contact Ion Science...

Страница 35: ...nts Batteries 4 x 1 5 V AA alkaline Time operable with batteries 30 hr occasional 40 hr continuous Monitoring Environmental operating conditions Ambient temperature range 15 F to 120 F 10 C to 50 C Te...

Страница 36: ...cm 2 s Resolution 2 pl cm 2 s hydrogen emanating from steel Linearity Linear over full range Reproducibility 5 or 1 pl cm 2 s whichever is greater Rate of response FSD 1 50 90 27 inch 70 cm conduit 8...

Страница 37: ...the prompt start up mode is displayed press D Display data log summary 10 s D DD Data download instructions 10 s D DDD Data download 30 s D DDDD Cancel download and return to Start up mode none Also d...

Страница 38: ...a log to 10 min data intervals Select D to end logging inde finite D low battery log full D DS Access log statistics file length maximum flux slope of present flux end log mode and return to operating...

Страница 39: ...T S Probes 13 01 09 V2 38 V3 25 Hydrosteel 6000 to V1 9 Warranty added on cover Declaration of Conformity updated Contents updated Warranty and Service info added pg 31 23 07 10 V2 38 V3 25 Hydrosteel...

Отзывы: