Gasclam
Ion Science Ltd
Page 16 of 37
Software
Different errors require different solutions, these are summarised below;
Error
Reason
Solution
Low Battery
Battery voltage is below 1.8V for 10s
GasClam will run but batteries need to
be changed
Flat Battery
Battery voltage is below 1.8V for 60s
The GasClam will not sample, batteries
must be changed
Filter Error
The moisture stripping filter is clogged
Change the moisture stripping filter.
If this does not work then call the
service centre
Immersion
Water in the borehole has risen to the
base of the GasClam
When water level drops the GasClam
will start sampling again
AD Overflow
The AD convertor is not working.
Check to see if it has happened only
once or if it is a permanent error. If
permanent reset the GasClam by
removing and replacing batteries. If
problem persists call the service
centre.
Flash Busy
The memory has timed out
Check to see if it has happened only
once or if it is a permanent error. If
permanent reset the GasClam by
removing and replacing batteries. If
problem persists call the service
centre.
Full Memory
The memory is full
Download GasClam and clear memory
Error SPI Flash
The flash memory has not worked
correctly
Check to see if it has happened only
once or if it is a permanent error. If
permanent reset the GasClam by
removing and replacing batteries. If
problem persists call the service
centre.
Error Write
Error writing to the flash memory
Check to see if it has happened only
once or if it is a permanent error. If
permanent reset the GasClam by
removing and replacing batteries. If
problem persists call the service
centre.
Pump
Pump is not working correctly
Call service centre.
After the problem has been resolved clear the error message by clicking „clr‟.
To help trouble shoot errors the service centre will need the .gcl, .gce and .csv file (see page 23)
Содержание GasClam
Страница 1: ...GasClam Instrument User Manual V1 5...
Страница 2: ...Gasclam Ion Science Ltd Page 1 of 37...
Страница 10: ...Gasclam Ion Science Ltd Page 9 of 37 Getting Started How the GasClam Works and access areas...
Страница 26: ...Gasclam Ion Science Ltd Page 25 of 37 Software Displayed multiple parameters...
Страница 31: ...Gasclam Ion Science Ltd Page 30 of 37 Other tests This section is for service centres...
Страница 35: ...Gasclam Ion Science Ltd Page 34 of 37 Service and Calibration Spare parts for users...
Страница 38: ...Gasclam Ion Science Ltd Page 37 of 37...