background image

Gasclam 

                                                                 

 

Ion Science Ltd 

Page 13 of 37 

Software    

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Main Screen  

 
Battery status – displays battery voltage alongside a power bar, the bar changes colour according to capacity providing a 

rough guide to battery life*. 
 
Green:  capacity is fine for a long sampling period 

 
Orange:  Batteries need to be replaced very soon 

 
Red:  Replace batteries immediately.  
 

*The voltage in the battery will decrease when the GasClam is sampling due to a load on the batteries, this is the voltage 
that should be used to assess battery condition (This is also the voltage that is logged). 

 
Note.  If the battery capacity is no longer sufficient for the running of the unit during a programme the unit automatically 
interrupts the cycle and switches to sleeping mode, this will be logged as a battery error. The data is stored in the flash 

memory therefore the data will remain in the GasClam even if the batteries are completely flat.  
 

Samples left –displays the remaining number of samples that can be stored in the memory.  
 

Samples taken/count – When the GasClam is in sleeping mode this reads „samples taken‟ indicating how many sampling 
points  are  stored  on  the  memory.    When  the  GasClam  is  in  sampling  mode  it  reads  „Sample  count‟  and  displays  the 
number  of  finished  cycles  and  the  total  number  of  cycles  in  the  programme.  For  example,  254/4500  means  that  254 

cycles from 4500 required cycles have finished.   
 

Sampling every - shows the period between sampling. For example, value “1 hour.” means that sampling frequency is 1 
hour. 
 

Finish on – displays the date and time that the sampling programme will finish. 
 

Serial number – displays the serial number of the unit 
 

COM port – displays which port the unit is connected to. 
 
Status – displays the mode of the unit: 

 
When the unit is running a programme the following modes are possible: 

 

Sampling – the GasClam is actively making a measurement. 

 

Measuring – the GasClam is between sampling periods. 

 

When the GasClam is not running a programme the following modes are possible: 

 

Sleeping – the unit is not running a programme. In this mode data can be downloaded and the unit programmed. 

 

Clear flash – the unit is erasing flash memory data 

 

If the GasClam is not functioning correctly the status will read 

„undefined status,‟ if this occurs contact customer services 

immediately. 
 
 
 

 
 

Содержание GasClam

Страница 1: ...GasClam Instrument User Manual V1 5...

Страница 2: ...Gasclam Ion Science Ltd Page 1 of 37...

Страница 3: ...nit 17 Set Up 18 Sampling Rate 18 Sampling Count 18 Device ID 18 Venting 19 Water Height 19 Unit Date and Time 20 Write Setting to GasClam 20 Erasing Data Memory 20 Returning to Main Screen 20 Downloa...

Страница 4: ...mp and filter assembly are located at the bottom of the housing Authorised Subject FTZ AO 210 NB 1026 Pikartsk 7 716 07 OSTRAVA RADVANICE I O 00577880 Type of Protection II 2G Ex d ib ib IIB T4 Certif...

Страница 5: ...nalysis in Excel Whilst connected to the software the settings of the GasClam unit can be altered including the frequency at which samples are taken and selecting venting options In addition to the se...

Страница 6: ...successfully removed concentration readings will be affected If the moisture stripping filter is completely saturated it can dissolve and it will then be pumped in to other sections of the GasClam res...

Страница 7: ...ce If chamber is known to flood or is already full of water Chose another chamber Or replace snorkel with bolt Effects on GasClam If submerged for an extended period the vent pipe may become flooded T...

Страница 8: ...l filter 2 O ring part 22 5 O ring part 21 5 Turning on the GasClam The GasClam can be switched on and off using the remote The remote connects to the communication port on top of the GasClam To start...

Страница 9: ...tion port 2 Pressure transducer port 3 Pressure transducer port cap 4 Gas inlet 5 Gas Outlet 6 Water proximity sensor 7 Pressure transducer hook 8 Battery compartment lid 9 Vent with hose barb 10 Rubb...

Страница 10: ...Gasclam Ion Science Ltd Page 9 of 37 Getting Started How the GasClam Works and access areas...

Страница 11: ...alternatively it can be downloaded from the Microsoft website Running the installation software Insert the installation CD or USB stick and copy the software folder on to the PC There are two installa...

Страница 12: ...sed by removing the 4 screws from the battery compartment lid see diagram To remove the batteries tilt the GasClam until they fall out Replace with the stipulated batteries positive terminal facing do...

Страница 13: ...nit before or after starting the software If the software is opened before the GasClam is attached the screen appears as below In this mode the options available are to view data see later or to close...

Страница 14: ...in sleeping mode this reads samples taken indicating how many sampling points are stored on the memory When the GasClam is in sampling mode it reads Sample count and displays the number of finished c...

Страница 15: ...ioning correctly Also this will remind the user that the GasClam has been started before leaving it for a period of unmanned data collection Firmware shows the firmware version in the unit Always use...

Страница 16: ...Gasclam Ion Science Ltd Page 15 of 37 Software Errors If the GasClam is not functioning correctly an error message will be displayed at the top of the Software home screen see below...

Страница 17: ...persists call the service centre Flash Busy The memory has timed out Check to see if it has happened only once or if it is a permanent error If permanent reset the GasClam by removing and replacing ba...

Страница 18: ...ton this reduces the discharge from the batteries When the off button is clicked a message will appear to ask if you really want to switch off If the Yes button is clicked the following appears The Ga...

Страница 19: ...r defined by the user in the Sample count box The maximum figure displayed is 65000 total memory space minus the number of stored samples the user defined value can be up to and including this The est...

Страница 20: ...ocated in a wet area and the vent blank is installed tick the vent plug inserted box Water height The GasClam can measure the water level in the borehole using an optional pressure transducer To enabl...

Страница 21: ...d to Write Settings to GasClam To activate a change in the setup up menu the write settings to GasClam button must be clicked If this is successful a message saying Settings written successfully appea...

Страница 22: ...le name and location the download begins when save is clicked the length of the download will depend upon the amount of data Three files are produced GCL This is the file format to read within the Gas...

Страница 23: ...n the main screen Selecting file for display To open a file click either the or the open button this allows you to browse through your computers folders Data Display Options There are 5 types of data...

Страница 24: ...Differential pressure between borehole and atmosphere If the value is negative it means the pressure in the borehole is lower than atmospheric and if the pressure is positive it is higher than atmosph...

Страница 25: ...e Graph Data view toggle button This opens the Graph view window shown on next page Selecting data channel To select a data channel click the desired parameter from the Data Channel box see below Mult...

Страница 26: ...Gasclam Ion Science Ltd Page 25 of 37 Software Displayed multiple parameters...

Страница 27: ...Import data into External Spreadsheet Data is provided as a CSV file and can be imported into various spreadsheets for manipulation This is normally done using the import function and selecting delim...

Страница 28: ...licking the view value box These values will update as the cursor is dragged over the graph Start Stop data To display the sampling log of the unit click the Start Stop Data button Information regardi...

Страница 29: ...this will be left in the code column Calibration Data In this section the calibration date and constants are displayed Error Messages All errors are logged and can be viewed here see pg 16 for detail...

Страница 30: ...to the sensor name then enter the concentration of the test gas in the Enter Values column If you want to test atmospheric gas the Enter values for Air button can be used to set the concentrations Bef...

Страница 31: ...Gasclam Ion Science Ltd Page 30 of 37 Other tests This section is for service centres...

Страница 32: ...owing the diagram below It is important to have the stand pipe a maximum of 3 cm above the base of the headworks this will ensure the GasClam will fit under the lid of the cover The headworks must be...

Страница 33: ...y explosive areas therefore it should only be serviced by qualified engineers Failure to do so will invalidate the warranty User serviceable parts Moisture stripping filter It is recommended that the...

Страница 34: ...Cable 4 25031 Push Button Control 5 25032 Moisture stripping filter tube 6 25033 Rubber Collar 8 10 25035 2 SA5 8 Vent hose barb vent blank 7 25034 Moisture Stripping Filter pack of 10 9 25036 Water...

Страница 35: ...Gasclam Ion Science Ltd Page 34 of 37 Service and Calibration Spare parts for users...

Страница 36: ...ric 800 1200 mBar 1 mBar Borehole Pressure Piezoelectric 800 1200 mBar 1 mBar Temperature Internal Chip 5 C to 50 C or 41 F to 122 F 1 C or 1 F Water depth Piezoelectric 0 25 m 0 01m Optional Memory 6...

Страница 37: ...aph image updated and general Layout of Manual updated Updated to version 1 3 06 06 09 06 04 75 3 7 8 GasClam v1 3 Extra sections added for 5 chamber GasClam Section how to protect the GasClam from wa...

Страница 38: ...Gasclam Ion Science Ltd Page 37 of 37...

Страница 39: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...