6.0

HANDGUN BUTTONS:

SQUARE:

 This has two functions: 

MEASURE

 MODE and 

SEARCH

 (bar graph) MODE.

On power-up the SEARCH mode is automatically selected, the touch screen displays a
bar graph. This is ideal for fast and ultra low-level leak detection (~10 E-8cc/sec). In this
mode high accuracy is not possible, just leak detection and location.

When a leak is discovered, the audible alarm sounds and the bar graph as well as the
handgun analogue gauge responds, usually full scale.

To change from SEARCH to MEASURE mode, press and hold the SQUARE button for 2
seconds, the RED light comes on and the display changes to 0.00. You are now in the
MEASURE high accuracy mode. If you have selected ‘Peak Hold Console’ leaks will be
measured and held ready for printing if required.

Pressing SQUARE once gives desensitise x1. The second press gives desensitise x2, as
pre set in USER SET-UP section in touch-screen. The LED marked SPECIAL flashes
slowly with x1, and faster with x2.

CIRCLE:

 This zeroes and cancels desensitise and may be used at any time. During the

short zero time the lights change from RED to GREEN. Do not leak test during this time.

During normal use it is recommended to press the CIRCLE key occasionally to
zero the detector to ambient conditions.

TRIANGLE:

 The RESULT STORE key. The peak reading shown on the touch-screen is

transferred to memory for later printout.

6.1:

TOP CAPS: 

A choice of top caps are available, Standard (P1-100-0003),

extended (P1-100-0004), and extended with a Stainless Steel tip (P1-100-0005). This
has a 5mm diameter and is designed for leak searching in enclosed spaces or where
rough (welded) surfaces must be checked.

6.2

FILTERS: 

It is essential that FILTER (P1-100-0005) is always fitted. This

provides a small micron filter and provides a seal for the secondary sinter. This sinter
forms part of the detector vacuum cell and is critical to good stable and accurate
operation. The flow down the top cap past the filter and into the vacuum chamber is
constantly monitored. The vacuum automatically compensates for filter restriction, to a
point when accuracy is threatened. At this time the LOW FLOW ERROR is displayed and
the filter should be changed.

All these items are available ex stock from ISL or your local agent.

Содержание GAS-CHECK SF6

Страница 1: ...INSTRUCTION MANUAL GAS CHECK SF6 AN INSTRUMENT FOR THE LOCATION AND MEASUREMENT OF LEAKS IN SF6 FILLED ELECTRICAL SWITCHGEAR User Manual iss 01 Software Version V1 39BB...

Страница 2: ...ion 6 7 3 Low flow error 7 7 4 Flare stall 7 7 3 Low battery voltage 7 7 5 Switching OFF 7 8 0 Switching OFF in case of fault 7 8 1 User setup 8 8 2 Set volume Contrast 8 9 0 Open Operator Level 8 9 1...

Страница 3: ...p after fuse has been removed 15 16 3 Pre set passwords on delivery 15 16 4 Settings on despatch 16 16 5 Citizen printer 16 16 6 Spares and consummables 16 3 0 GENERAL This manual is addressed to thos...

Страница 4: ...contain a microchip programmed with specific sensor information At Switch ON the console automatically reads sensor information Smart Sensors are not radioactive and contain nothing harmful Life is ap...

Страница 5: ...may confidently measure This is OFF whenever the instrument is in standby or programming mode error condition or the actual reading is above 20 of the alarm level Red 2 NOT READY This light is on whe...

Страница 6: ...used at any time During the short zero time the lights change from RED to GREEN Do not leak test during this time During normal use it is recommended to press the CIRCLE key occasionally to zero the d...

Страница 7: ...for you press Exit Should this not be the case please refer to paragraphs Possible Fault Conditions During Operation and Switching Off in Case of Fault in this manual 7 1 MEASURING OPERATION The GAS C...

Страница 8: ...entration of SF6 many times above the measuring range 2 High vacuum in the probe caused by a partially blocked or aged sensor The instrument will attempt to compensate for these until a level is reach...

Страница 9: ...the present value of a parameter will already be entered You may then confirm or delete and re enter 8 2 Set volume Contrast Using the three slider type controls on this screen select contrast for be...

Страница 10: ...vel alerts the ALARM LED in the handgun Only the alarm level corresponding to the chosen mode ml sec or ppm can be entered Valid alarm levels are for leak rates 0 5 E 6 5 E 7 ml s to 20 E 6 200 E 7 ml...

Страница 11: ...be assigned for 6 different operators please refer to Changing Passwords section 10 4 10 3 Access codes 10 4 CHANGING PASSWORDS Use one of the eight buttons to select the password you wish to change t...

Страница 12: ...Air is measured first which takes 4 seconds A time bar is displayed along with an audible bleep The display then instructs the operator to put the probe to the reference leak When this is done the fi...

Страница 13: ...when the STORE key is pressed Pressing either the STORE or CLEAR buttons resets this 12 3 Display units The touch screen display and alarm level entries as well as reference leak values will operate...

Страница 14: ...e following will occur The touch screen screen contrast sensor and sample flow regulation are switched off Vacuum is released Lifetime of all major components is saved If POWER DOWN is selected the in...

Страница 15: ...r operation The main function of the filter is to block out short wave noise Filter Count represents the number of measuring cycles over which the signal must be continuously above or below the averag...

Страница 16: ...ew sensor and sets these values in the console memory this may take 10 50 seconds 15 2 Changing the Probe To change the sensor You Must Power Down Never unplug the probe while the red lights on the co...

Страница 17: ...er set up ALL 4 SWITCHES MUST BE IN THE OFF POSITION Switch the printer ON using the side mounted slide switch The POWER and SEL green lights should be lit Press SEL to turn OFF the light and use LF t...

Отзывы: