ion science GAS-CHECK SF6 Скачать руководство пользователя страница 17

Page 16

16.4 

Settings on despatch as found in User Setup

Instrument serial Number P1

Date

Peak Hold (gauge)

0.5 s

Peak Hold (screen)

AUX DISP

Display Units

E-6 ml/s

Desen. Factor 1

5 x

Desen  Factor 2

10 x

Standby Mode

Light + sensor

Standby Delay

15 min

RS232 Protocol

DTR/DSR

Sound Type

HARMONIC

Sound begins at 

20% Alarm

Live tick

OFF

Filter Threshold

5

Filter Count

10

16.5

A Citizen dot matrix printer type iDP together with a mains transformer is supplied as an
option. Additional printer rolls can be obtained from ISL or from your local computer
printer supplier. Below the paper roll is a DIL switch having 4 switches. With DTR/DRS
selected from the user set-up ALL 4 SWITCHES MUST BE IN THE OFF POSITION.

Switch the printer ON using the side mounted slide switch. The POWER and SEL green
lights should be lit. Press SEL to turn OFF the light and use LF to advance the paper.
Now switch SEL ON and you can now print the stored results by tapping the screen,
entering OPEN OPERATOR LEVEL then DATA PRINTOUT. The printer will then
operate.

Note that if you try to print with SEL OFF the screen will show print in progress, and then
time out.

If you have entered your operator or company information in Operator ID see 9.1 page 8
This will appear in the printed message.

16.6 Spares & Consumables:

Description                                     Part No
Sensor probe complete.

P1 100 0000

Steel tipped top-cap (25mm)

P1 100 0004

supplied in packs of 3

Filter

P1 100 0005

supplied in packs of 10

Manual Iss 07
Software version V1.39BB
Compiled by Mel Vinton.

Содержание GAS-CHECK SF6

Страница 1: ...INSTRUCTION MANUAL GAS CHECK SF6 AN INSTRUMENT FOR THE LOCATION AND MEASUREMENT OF LEAKS IN SF6 FILLED ELECTRICAL SWITCHGEAR User Manual iss 01 Software Version V1 39BB...

Страница 2: ...ion 6 7 3 Low flow error 7 7 4 Flare stall 7 7 3 Low battery voltage 7 7 5 Switching OFF 7 8 0 Switching OFF in case of fault 7 8 1 User setup 8 8 2 Set volume Contrast 8 9 0 Open Operator Level 8 9 1...

Страница 3: ...p after fuse has been removed 15 16 3 Pre set passwords on delivery 15 16 4 Settings on despatch 16 16 5 Citizen printer 16 16 6 Spares and consummables 16 3 0 GENERAL This manual is addressed to thos...

Страница 4: ...contain a microchip programmed with specific sensor information At Switch ON the console automatically reads sensor information Smart Sensors are not radioactive and contain nothing harmful Life is ap...

Страница 5: ...may confidently measure This is OFF whenever the instrument is in standby or programming mode error condition or the actual reading is above 20 of the alarm level Red 2 NOT READY This light is on whe...

Страница 6: ...used at any time During the short zero time the lights change from RED to GREEN Do not leak test during this time During normal use it is recommended to press the CIRCLE key occasionally to zero the d...

Страница 7: ...for you press Exit Should this not be the case please refer to paragraphs Possible Fault Conditions During Operation and Switching Off in Case of Fault in this manual 7 1 MEASURING OPERATION The GAS C...

Страница 8: ...entration of SF6 many times above the measuring range 2 High vacuum in the probe caused by a partially blocked or aged sensor The instrument will attempt to compensate for these until a level is reach...

Страница 9: ...the present value of a parameter will already be entered You may then confirm or delete and re enter 8 2 Set volume Contrast Using the three slider type controls on this screen select contrast for be...

Страница 10: ...vel alerts the ALARM LED in the handgun Only the alarm level corresponding to the chosen mode ml sec or ppm can be entered Valid alarm levels are for leak rates 0 5 E 6 5 E 7 ml s to 20 E 6 200 E 7 ml...

Страница 11: ...be assigned for 6 different operators please refer to Changing Passwords section 10 4 10 3 Access codes 10 4 CHANGING PASSWORDS Use one of the eight buttons to select the password you wish to change t...

Страница 12: ...Air is measured first which takes 4 seconds A time bar is displayed along with an audible bleep The display then instructs the operator to put the probe to the reference leak When this is done the fi...

Страница 13: ...when the STORE key is pressed Pressing either the STORE or CLEAR buttons resets this 12 3 Display units The touch screen display and alarm level entries as well as reference leak values will operate...

Страница 14: ...e following will occur The touch screen screen contrast sensor and sample flow regulation are switched off Vacuum is released Lifetime of all major components is saved If POWER DOWN is selected the in...

Страница 15: ...r operation The main function of the filter is to block out short wave noise Filter Count represents the number of measuring cycles over which the signal must be continuously above or below the averag...

Страница 16: ...ew sensor and sets these values in the console memory this may take 10 50 seconds 15 2 Changing the Probe To change the sensor You Must Power Down Never unplug the probe while the red lights on the co...

Страница 17: ...er set up ALL 4 SWITCHES MUST BE IN THE OFF POSITION Switch the printer ON using the side mounted slide switch The POWER and SEL green lights should be lit Press SEL to turn OFF the light and use LF t...

Отзывы: