![INW PT2X Скачать руководство пользователя страница 26](http://html.mh-extra.com/html/inw/pt2x/pt2x_instruction-manual_2085811026.webp)
INW - PT2X Instruction Manual
25
Staff Gauge
When using the Staff Gauge setting, you may want to change the channel label at the
top of the Settings window to read “Staff Gauge”.
NOTE: Be sure to use the same units for all measurements on this page. For example,
if you have the program set to display in Ft H2O then be sure to enter the barometric
pressure and staff gauge readings in Ft H2O.
-- Computing Offset --
• If using an absolute sensor, enter the current barometric pressure in the
Barometric
box.
• Place the sensor in the well.
• Note the current level on your staff gauge.
• Enter this value in the
Staff
box.
• Click
the
Measur
e button.
• When readings have stabilized, click the
Accept
button in the pop-up box.
-- Applying Offset --
• Click
the
Apply
button to apply the calculated offset.
• The calculated offset will be transferred to the offset fi eld near the top of the
window.
•
Click OK to save the values to the sensor!!!
-- Verifying Settings --
• Using the Real Time Monitor, take a few readings to verify it is reading as
expected.
IMPORTANT NOTE!
When taking readings on an
absolute
sensor, you will need to
SUBRACT
the current barometric pressure to get a staff gauge reading. See the
Barometric Compensation Utility Application Note (9C0275) for information
on how the Barometric Utility can help automate this process.