INW - PT2X Instruction Manual
Appendix B: Settings and Calibration
Settings and calibration values can only be changed when there is no data on the sensor.
Be sure to retrieve any data and then erase the data on the sensor before proceeding.
The temperature channel rarely needs adjustment. If you think your temperature
channel needs calibrating, contact your service representative.
Before leaving the factory, your PT2X has been inspected using precision instruments.
However, you may want to change some of the settings for the pressure channel for the
following reasons:
• To set a specifi c zero-reference point for submergence
• To return values as depth-to-water
• To return values as elevation in relation to sea-level
• To return values relative to a staff gauge
Preparation
• Run Aqua4Plus Lite or Aqua4Plus and scan for sensors.
• On the Sensor Map, click the sensor you want to adjust.
• From
Aqua4Plus
Lite:
o Select the units you want to use for measurements, i.e., psi, feet of
water, inches of mercury, etc. Set this from the Confi gure Menu |
Program
Confi guration | Set Computer Display Units.
o Open the Settings and Calibration window by clicking on the
Settings button .
• From
Aqua4Plus:
o Select the units you want to use for measurements, i.e., psi, feet of
water, inches of mercury, etc. Set this from the Options Menu | Display
Units.
o Open the Settings and Calibration window from the Confi gure Menu |
Settings and Calibration.
Select the setting type from the Select Setting Type dropdown box. A picture will
display illustrating the type of selection. Step-by-step instructions will be displayed on
the right.